Chapter 6
First Light
55
19. When using WinSpec/32, go to the Experiment Setup
►
ROI Setup tab and select
Imaging Mode. See
Figure 6-7: Typical WinX/32 Experiment Setup
►
ROI Setup Tab Dialog
20. Click OK to save all parameter settings and dismiss the dialog.
21. Verify that room or ambient lighting is subdued.
22. Set the PI-MAX3 I.I.T switch to on.
NOTE:
The PI-MAX3 overload alarm may sound briefly and then
stop. This is normal and is not a cause for concern.
However, if the alarm sounds continuously, even when
there is no light entering the camera/spectrograph, there is
something wrong. Turn off all power and contact Teledyne
Princeton Instruments Customer Service for assistance.
Refer to
on page 248 for complete
information.
4
4
11-
01
29
_
0020
Summary of Contents for Princeton Instruments PI-MAX 3
Page 1: ...www princetoninstruments com PI MAX 3 System Manual 4411 0129 Issue 2 August 20 2019 ...
Page 32: ...32 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 38: ...38 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 124: ...124 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 150: ...150 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 174: ...174 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 178: ...178 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 190: ...190 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 214: ...214 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 236: ...236 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 244: ...244 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 249: ...This page is intentionally blank ...