Cooling
Cooling within the PI-MAX3 is performed by a cooling fan and a multi-stage Peltier
cooler that is thermally coupled to the CCD.
A ventilation fan runs continuously to remove heat generated by the thermoelectric
cooler and the electronics. Air is drawn into the camera through the rear grill, picks up
the heat from the electronics and the cooler, and is then exhausted through the side
ventilation slots.
With air-cooling alone, at an ambient temperature of 25°C, temperature lock at –25°C
will generally occur within ten to fifteen minutes. Note that the exact cooling
performance is a function of the CCD array installed. Also, if the lab is particularly warm,
achieving temperature lock might take longer or not occur at all.
NOTE:
Liquid coolant circulation can also be used with the
PI-MAX3 camera.
Photocathode cooling to reduce equivalent background illumination (EBI) can be
achieved via a dry nitrogen source.
NOTE:
Clear PVC tubing (3 feet, 5/32” OD, 1/32” wall,
McMaster-Carr 5006K42) is supplied with the PI-MAX3 to
connect to a dry nitrogen source for cooling the
photocathode.
2.1.3
GigE Ethernet Card
2.1.4
Extender Bracket Kit
Each PI-MAX3 camera is shipped with an Extender Bracket kit that, when mounted to
the rear of the camera, allows the PI-MAX3 to be secured to a laboratory table with
tapped mounting hole arrays with ¼-20 on 1” spacing, or M6 on 25 mm spacing.
4411-0129_0003
This card must be installed in the host computer to
control the PI-MAX3 camera and to receive data from
the camera.
Summary of Contents for Princeton Instruments PI-MAX 3
Page 1: ...www princetoninstruments com PI MAX 3 System Manual 4411 0129 Issue 2 August 20 2019 ...
Page 32: ...32 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 38: ...38 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 124: ...124 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 150: ...150 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 174: ...174 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 178: ...178 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 190: ...190 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 214: ...214 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 236: ...236 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 244: ...244 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 249: ...This page is intentionally blank ...