Chapter 9
LightField and Gated Operation
145
The sequence of operations is similar to the Sequential experiments. After focusing the
camera on the fluorescing sample, the appropriate Gain is selected. See
Figure 9-22: Typical Common Acquisition Settings Expander: Configure Gain
Gate width and gate delay are set in such a way that the intensifier is gated ON during
the entire event (in this case the event is a 60 ns fluorescence). See
.
Figure 9-23: Repetitive Gating Setup: 100 ns Width, 25 ns Delay
44
11
-0
12
9_
0
0
83
4
4
11-
012
9
_0
09
3
Summary of Contents for Princeton Instruments PI-MAX 3
Page 1: ...www princetoninstruments com PI MAX 3 System Manual 4411 0129 Issue 2 August 20 2019 ...
Page 32: ...32 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 38: ...38 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 124: ...124 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 150: ...150 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 174: ...174 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 178: ...178 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 190: ...190 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 214: ...214 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 236: ...236 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 244: ...244 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 249: ...This page is intentionally blank ...