![Teledyne Princeton Instruments PI-MAX 3 System Manual Download Page 58](http://html.mh-extra.com/html/teledyne/princeton-instruments-pi-max-3/princeton-instruments-pi-max-3_system-manual_1080556058.webp)
58
PI-MAX
®
3 System Manual
Issue 2
6.2
LightField First Light
This section provides the step-by-step procedure for acquiring an imaging
measurement in LightField for the first time. The purpose of this section is to gain
familiarity with the operation of the system and to show that it is functioning properly.
Once basic familiarity has been established, additional, more complex configurations
can be implemented.
6.2.1
Required Equipment
The following items are required to operate the PI-MAX3 camera system in accordance
with the procedure in this section:
•
Teledyne Princeton Instruments PI-MAX3 camera with C-mount adapter;
•
PI-MAX3 power supply;
•
User-supplied C-mount lens with smallest aperture of f/16 or f/22;
•
Host Computer equipped with a GigE Ethernet interface card;
•
Standard Ethernet cable;
•
LightField Application Software.
6.2.2
Procedure
This procedure assumes:
•
The PI-MAX3 system has been set up according to prior chapters;
•
Prior sections of this chapter have been read;
•
Familiarity with the LightField software;
•
The PI-MAX3 camera is being operated in imaging mode;
•
The target that is being used is a sharp image, text, or a drawing that can be
used to verify that the PI-MAX3 is seeing and can be used to maximize focus.
NOTE:
This procedure is based on LightField. Basic familiarity with
the LightField software is assumed. If this is not the case,
you may want to refer to the online help system while
performing this procedure.
Perform the following procedure to configure the PI-MAX3 system to acquire an image
with LightField for the first time:
WARNING!
All system components must be turned off before
connecting or disconnecting cables.
1. Connect the PI-MAX3 power supply to:
•
The rear of the PI-MAX3;
•
An AC power source.
2. Connect one end of the GigE cable (P/N: 6050-0621) to the PI-MAX3 GigE
connector and the other end to the host computer's GigE Interface card connector.
3. Set the PI-MAX3 I.I.T. On/Off switch to off
4. Confirm that only room light will be present.
5. If a C-mount adapter is not on the camera, refer to
, on page 223 for complete information about
installing it.
Summary of Contents for Princeton Instruments PI-MAX 3
Page 1: ...www princetoninstruments com PI MAX 3 System Manual 4411 0129 Issue 2 August 20 2019 ...
Page 32: ...32 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 38: ...38 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 124: ...124 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 150: ...150 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 174: ...174 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 178: ...178 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 190: ...190 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 214: ...214 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 236: ...236 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 244: ...244 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 249: ...This page is intentionally blank ...