Chapter 9
LightField and Gated Operation
137
b. On the Analog to Digital Conversion expander, select the desired Speed and
Analog Gain. See
Figure 9-10: Typical Analog to Digital Conversion Expander
c. On the Trigger expander, select Internal. See
.
Figure 9-11: Typical Trigger Expander: Internal Trigger Source Selected
7. After setting the parameters and making sure the ambient light level is low, click on
the View tab, followed by Run to begin acquiring data.
NOTE:
This data will not be saved.
8. After you verify that the camera is seeing, stop data acquisition.
9. On the Region of Interest expander, select Full Sensor, Binned and enter 1024 in the
Bin H field. See
.
4
4
11-
012
9
_0
084
4
4
11-
0129
_0
085
Summary of Contents for Princeton Instruments PI-MAX 3
Page 1: ...www princetoninstruments com PI MAX 3 System Manual 4411 0129 Issue 2 August 20 2019 ...
Page 32: ...32 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 38: ...38 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 124: ...124 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 150: ...150 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 174: ...174 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 178: ...178 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 190: ...190 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 214: ...214 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 236: ...236 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 244: ...244 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 249: ...This page is intentionally blank ...