122
PI-MAX
®
3 System Manual
Issue 2
The sequence of operations is similar to the Sequential experiments. After focusing the
camera on the fluorescing sample, the camera is set in Gate mode and the appropriate
Gain is selected. See
Figure 8-29: Typical Experiment Setup Dialog: Main Tab, Gain Configuration
Gate width and gate delay are set in such a way that the intensifier is gated ON during
the entire event (in this case the event is a 60 ns fluorescence). See
.
Figure 8-30: Repetitive Gating Setup: 100 ns Width, 10 ns Delay
4
4
11-
012
9
_0
06
0
4
4
11-
0129
_00
77
Summary of Contents for Princeton Instruments PI-MAX 3
Page 1: ...www princetoninstruments com PI MAX 3 System Manual 4411 0129 Issue 2 August 20 2019 ...
Page 32: ...32 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 38: ...38 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 124: ...124 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 150: ...150 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 174: ...174 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 178: ...178 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 190: ...190 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 214: ...214 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 236: ...236 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 244: ...244 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 249: ...This page is intentionally blank ...