BLZZRD
™
Portable Refrigerated Sampler
Section 2 Installation/Preparation
2-5
To inspect the pump tube:
CAUTION
Moving parts can cause injuries. Remove power before
inspecting pump tube.
1. Disconnect power from the sampler.
2. Unlatch the pump band and swing it away to expose the
pump tube. (The band is the rounded metal cover that
holds the tube in the pump.)
3. Visually inspect the pump tube for cracks where it is com-
pressed by the rollers. If the tube is cracked, it must be
replaced. Section 9.4 provides detailed instructions on
replacing the pump tube and resetting the pump tube
counter.
4. Visually inspect the inside of the pump housing. The hous-
ing, drain holes, and rollers should be free from debris.
Clean if necessary.
5. Latch the pump band.
2.5 Installing the Bottle
Kit
The sampler is shipped from the factory with sample bottles in
place. When using the sampler for the first time, you will only
need to remove the bottle lids. When installing cleaned bottles or
a new kit, keep these guidelines in mind:
•
Install all bottles to ensure that they remain in place.
•
Align the bottles correctly. If the bottles are misaligned,
the sampler may miss the bottle mouth, or the sampler
will deposit samples in the “wrong” bottle.
•
Each time you change bottle configurations, check the
program settings for the number of bottles and bottle
volume.
2.5.1 Installing the 14-Bottle
Kit
To install the 14-Bottle Kit (68-2970-001):
1. Attach the distributor arm to the distributor shaft under-
neath the sampler controller. Note that the metal shaft and
arm are keyed to ensure correct alignment. Secure the dis-
tributor arm with the knurled nut.
2. Refer to Figure 2-2. Attach the 13 inch (33 cm) discharge
tube (Teledyne ISCO Part Number 60-2973-054) to the
bulkhead fitting. Route the tube through the stain-
less-steel tube holder.
3. Route the other end of the tube through the spring on the
distributor arm. Then, place the end of the discharge tube
through the opening in the end of the distributor arm. The
end of the tube should protrude below the arm
1
/
16
to
1
/
8
inch (see Figure 2-2). If too much tube protrudes from the
arm, it can catch on bottles and jam the distributor arm.
Summary of Contents for Everywhereyoulook BLZZRD
Page 2: ......
Page 4: ......
Page 8: ...BLZZRD Portable Refrigerated Sampler Safety vi ...
Page 16: ...BLZZRD Portable Refrigerated Sampler Table of Contents xiv ...
Page 34: ...BLZZRD Portable Refrigerated Sampler Section 1 Introduction 1 18 ...
Page 58: ...BLZZRD Portable Refrigerated Sampler Section 2 Installation Preparation 2 24 ...
Page 72: ...BLZZRD Portable Refrigerated Sampler Section 3 Programming Introduction 3 14 ...
Page 88: ...BLZZRD Portable Refrigerated Sampler Section 4 Standard Programming 4 16 ...
Page 219: ...BLZZRD Portable Refrigerated Sampler Appendix B Material Safety Data Sheets B 8 ...
Page 233: ...BLZZRD Portable Refrigerated Sampler Appendix D Replacement Parts D 6 ...
Page 241: ...BLZZRD Portable Refrigerated Sampler Appendix F Battery Selection Guide F 4 ...