BLZZRD Portable Refrigerated Sampler
Section 5 Extended Programming
5-14
In the
example, a rate of change is set so
that if the level rises more than one foot in a two hour time
period, the event condition would be enabled and a sample taken.
In determining if the rise occurs in this example, the sampler
looks back at the previous two hour time period to see if there is a
rise of more than one foot between any valley (low point) or peak
(high point) within that time period. Once the condition becomes
true (a rise of more than one foot is noted), the event condition is
enabled. It will not be enabled again unless the condition
becomes false, and then true again, within the two hour time
window.
In event pacing, up to three different rate of change event condi-
tions can be programmed (for two-part programs, a total of six
rate of change event conditions can be defined, three for each
part).
5.8.2 Nonuniform Time
Pacing
The extended time pacing options are uniform time and non-
uniform time. “Uniform time” pacing is identical to “time” pacing
in standard programming. Nonuniform time pacing uses an
irregular interval between sample events; each interval is indi-
vidually programmable. There are three types of nonuniform
time pacing:
•
Nonuniform clock time pacing.
•
Nonuniform interval pacing.
•
Random interval pacing.
5.8.3 Nonuniform Clock
Time Pacing
For clock time pacing, enter specific times for each sample event
(see screen 2 in the example). You can enter as many as 99 clock
times, but the interval between times cannot be greater than 24
hours. (Remember to use a 24-hour clock when entering times.)
The sampler takes a sample at the start time.
5.8.4 Nonuniform Interval
Pacing
For nonuniform interval pacing, enter the number of sample
events spaced at intervals of minutes: 12 samples at 5 minute
intervals, 6 samples at 10 minute intervals, and so on. The
sampler accepts as many as 99 nonuniform interval entries. The
sampler takes a sample at start time.
5.8.5 Random Interval
Pacing
To program the sampler for random interval pacing, enter the
length of time you want to sample. The run time is the only
random programming setting you’ll need to enter because the
sampler derives the number of sample events required for the
program from the distribution settings. (You can combine
random pacing with any distribution.) Each time you run the
program, the sampler generates a new set of random intervals.
This makes each sample event unpredictable from run to run.
In both nonuniform clock time pacing and nonuniform interval
pacing, the sampler takes a sample at the start time. For random
pacing, however, it takes the first sample at the end of the first
interval, not at the start time.
Summary of Contents for Everywhereyoulook BLZZRD
Page 2: ......
Page 4: ......
Page 8: ...BLZZRD Portable Refrigerated Sampler Safety vi ...
Page 16: ...BLZZRD Portable Refrigerated Sampler Table of Contents xiv ...
Page 34: ...BLZZRD Portable Refrigerated Sampler Section 1 Introduction 1 18 ...
Page 58: ...BLZZRD Portable Refrigerated Sampler Section 2 Installation Preparation 2 24 ...
Page 72: ...BLZZRD Portable Refrigerated Sampler Section 3 Programming Introduction 3 14 ...
Page 88: ...BLZZRD Portable Refrigerated Sampler Section 4 Standard Programming 4 16 ...
Page 219: ...BLZZRD Portable Refrigerated Sampler Appendix B Material Safety Data Sheets B 8 ...
Page 233: ...BLZZRD Portable Refrigerated Sampler Appendix D Replacement Parts D 6 ...
Page 241: ...BLZZRD Portable Refrigerated Sampler Appendix F Battery Selection Guide F 4 ...