APPENDIX A-2: Setup Variables For Serial I/O,
Revision B.3
Model 100E Instruction Manual
236
045150110 Rev B.3
Setup Variable
Numeric
Units
Default Value
Value Range
Description
TEST_CHAN_ID — NONE
NONE,
PMT
READING,
UV
READING,
VACUUM
PRESSURE
3
,
SAMPLE
PRESSURE,
SAMPLE
FLOW,
RCELL TEMP,
CHASSIS
TEMP,
IZS TEMP
1
,
PMT TEMP,
HVPS
VOLTAGE
Diagnostic analog output
ID. Enclose value in
double quotes (“) when
setting from the RS-232
interface.
REMOTE_CAL_MODE —
LOW
LOW,
HIGH
Range to calibrate during
contact-closure and
Hessen calibration.
Enclose value in double
quotes (“) when setting
from the RS-232
interface.
PASS_ENABLE
—
OFF
OFF, ON
ON enables passwords;
OFF disables them.
STABIL_FREQ
Seconds
10
1 to 300
Stability measurement
sampling period.
STABIL_SAMPLES
Samples
25
2 to 40
Number of samples in
concentration stability
reading.
RCELL_CYCLE
Seconds
2
0.5 to 30
Sample chamber
temperature control cycle
period.
RCELL_PROP
1/ºC
0.3 (prop. band =
3.3 ºC)
0 to 10
Sample chamber
temperature PID
proportional coefficient.
RCELL_INTEG
—
0.005
0 to 10
Sample chamber
temperature PID integral
coefficient.
RCELL_DERIV
—
0.5
0 to 10
Sample chamber
temperature PID
derivative coefficient.
IZS_CYCLE
1
Seconds
2
0.5 to 30
IZS temperature control
cycle period.
Summary of Contents for 100E
Page 2: ...Model 100E Instruction Manual ...
Page 4: ...Model 100E Instruction Manual ii 045150102 Rev XB1 ...
Page 12: ...Model 100E Instruction Manual x 045150102 Rev XB1 ...
Page 20: ...M100E Documentation Model 100E Instruction Manual 18 045150102 Rev XB1 User Notes ...
Page 26: ...Specifications Approvals and Warranty Model 100E Instruction Manual 24 045150102 Rev XB1 ...
Page 46: ...Getting Started Model 100E Instruction Manual 44 045150102 Rev XB1 User Notes ...
Page 128: ...Operating Instructions Model 100E Instruction Manual 126 045150102 Rev XB1 User Notes ...
Page 156: ...EPA Protocol Calibration Model 100E Instruction Manual 154 045150102 Rev XB1 ...