Model 100E Instruction Manual
Operating Instructions
045150102 Rev XB1
69
6.5.6.
Range Units
The M100E can display concentrations in parts per billion (10
9
mols per mol, PPB), parts
per million (10
6
mols per mol, PPM), micrograms per cubic meter (µg/m
3
, UG) or
milligrams per cubic meter (mg/m
3
, MG). Changing units affects all of the display, analog
outputs, COM port and iDAS values for all reporting ranges regardless of the analyzer’s
range mode.
To change the concentration units:
SAMPLE RANGE = 500.0 PPB NOX=X.X
< TST TST > CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK MORE
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET
UNIT
EXIT
SETUP X.X
CONC UNITS: PPM
PPM PPB
UGM
MGM
ENTER
EXIT
Select the preferred
concentration unit.
SETUP X.X
CONC UNITS: UGM
PPM PPB
UGM
MGM
ENTER EXIT
EXIT
returns
to the main menu.
ENTR
accepts
the new unit,
EXIT
returns
to the
SETUP
menu.
NOTE
Concentrations displayed in mg/m
3
and µg/m
3
use 0°C and 760 Torr as
standard temperature and pressure (STP). Consult your local regulations for
the STP used by your agency. Here are the conversion factors from volumetric
to mass units used in the M200E:
NO: ppb x 1.34 = µg/m
3
; ppm x 1.34 = mg/m
3
NO
2
: ppb x 2.05 = µg/m
3
; ppm x 2.05 = mg/m
3
Summary of Contents for 100E
Page 2: ...Model 100E Instruction Manual ...
Page 4: ...Model 100E Instruction Manual ii 045150102 Rev XB1 ...
Page 12: ...Model 100E Instruction Manual x 045150102 Rev XB1 ...
Page 20: ...M100E Documentation Model 100E Instruction Manual 18 045150102 Rev XB1 User Notes ...
Page 26: ...Specifications Approvals and Warranty Model 100E Instruction Manual 24 045150102 Rev XB1 ...
Page 46: ...Getting Started Model 100E Instruction Manual 44 045150102 Rev XB1 User Notes ...
Page 128: ...Operating Instructions Model 100E Instruction Manual 126 045150102 Rev XB1 User Notes ...
Page 156: ...EPA Protocol Calibration Model 100E Instruction Manual 154 045150102 Rev XB1 ...