Model 100E Instruction Manual
APPENDIX A-2: Setup Variables For Serial
I/O, Revision B.3
045150110 Rev B.3
233
Setup Variable
Numeric
Units
Default Value
Value Range
Description
VAC_SAMP_RATIO
3
—
0.53
0.1 to 2
Maximum vacuum
pressure / sample
pressure ratio for valid
sample flow calculation.
SAMP_PRESS_SET “Hg-A 29.92
Warnings:
15 to 35
0 to 100
Sample pressure set point
for pressure compensation
and warning limits.
SAMP_PRESS_SLOPE
—
1
0.5 to 1.5
Sample pressure slope
correction factor (adjusted
pressure = measured
pressure x slope).
VAC_PRESS_SET
3
“Hg
6
Warnings:
3 to 10
0 to 100
Vacuum pressure set point
for pressure compensation
and warning limits.
BOX_SET ºC
30
Warnings:
8 to 50
5 to 60
Box temperature warning
limits. Set point is not
used.
PMT_SET ºC
7
Warnings:
2 to 12
0 to 40
PMT temperature set point
and warning limits.
Summary of Contents for 100E
Page 2: ...Model 100E Instruction Manual ...
Page 4: ...Model 100E Instruction Manual ii 045150102 Rev XB1 ...
Page 12: ...Model 100E Instruction Manual x 045150102 Rev XB1 ...
Page 20: ...M100E Documentation Model 100E Instruction Manual 18 045150102 Rev XB1 User Notes ...
Page 26: ...Specifications Approvals and Warranty Model 100E Instruction Manual 24 045150102 Rev XB1 ...
Page 46: ...Getting Started Model 100E Instruction Manual 44 045150102 Rev XB1 User Notes ...
Page 128: ...Operating Instructions Model 100E Instruction Manual 126 045150102 Rev XB1 User Notes ...
Page 156: ...EPA Protocol Calibration Model 100E Instruction Manual 154 045150102 Rev XB1 ...