APPENDIX A-2: Setup Variables For Serial I/O,
Revision B.3
Model 100E Instruction Manual
230
045150110 Rev B.3
APPENDIX A-2: Setup Variables For Serial I/O, Revision B.3
Table A-1: M100E Setup Variables, Revision B.3
Setup Variable
Numeric
Units
Default Value
Value Range
Description
DAS_HOLD_OFF
Minutes
15
0.5 to 20
Duration of DAS hold off
period.
TPC_ENABLE — ON,
OFF
4
OFF, ON
ON enables temperature
and pressure
compensation; OFF
disables it.
RCELL_SET ºC
50
Warnings:
45 to 55
30 to 70
Sample chamber
temperature set point and
warning limits.
IZS_SET
1
ºC
50
Warnings:
45 to 55
30 to 70
IZS temperature set point
and warning limits.
DYN_ZERO
—
OFF
OFF, ON
ON enables contact
closure dynamic zero; OFF
disables it.
DYN_SPAN
—
OFF
OFF, ON
ON enables contact
closure dynamic span;
OFF disables it.
CONC_PRECISION
—
1
AUTO, 0, 1, 2,
3, & 4
Number of digits to
display to the right of the
decimal point for
concentrations on the
display. Enclose value in
double quotes (“) when
setting from the RS to 232
interface.
CLOCK_ADJ
Sec./Day
0
-60 to 60
Time-of-day clock speed
adjustment.
LANGUAGE_SELECT —
ENGL
ENGL,
SECD,
EXTN
Selects the language to
use for the user interface.
Enclose value in double
quotes (“) when setting
from the RS-232
interface.
MAINT_TIMEOUT
Hours
2
0.1 to 100
Time until automatically
switching out of software-
controlled maintenance
mode.
CONV_TIME
—
33 MS
33 MS,
66 MS,
133 MS,
266 MS,
533 MS,
1 SEC,
2 SEC
Conversion time for PMT
and UV detector channels.
Enclose value in double
quotes (“) when setting
from the RS-232
interface.
Summary of Contents for 100E
Page 2: ...Model 100E Instruction Manual ...
Page 4: ...Model 100E Instruction Manual ii 045150102 Rev XB1 ...
Page 12: ...Model 100E Instruction Manual x 045150102 Rev XB1 ...
Page 20: ...M100E Documentation Model 100E Instruction Manual 18 045150102 Rev XB1 User Notes ...
Page 26: ...Specifications Approvals and Warranty Model 100E Instruction Manual 24 045150102 Rev XB1 ...
Page 46: ...Getting Started Model 100E Instruction Manual 44 045150102 Rev XB1 User Notes ...
Page 128: ...Operating Instructions Model 100E Instruction Manual 126 045150102 Rev XB1 User Notes ...
Page 156: ...EPA Protocol Calibration Model 100E Instruction Manual 154 045150102 Rev XB1 ...