Model 100E Instruction Manual
APPENDIX A-3: Warnings and Test Functions,
Revision B.3
045150110 Rev B.3
241
Table A-3: M100E Test Functions, Revision B.3
TEST FUNCTION
MESSAGE TEXT
DESCRIPTION
RANGE RANGE=500.0
PPB
3
D/A range in single or auto-range modes.
RANGE1 RANGE1=500.0
PPB
3
D/A #1 range in independent range mode.
RANGE2 RANGE2=500.0
PPB
3
D/A #2 range in independent range mode.
STABILITY STABIL=0.0
PPB
3
Concentration stability (standard deviation based on
setting of
STABIL_FREQ
and
STABIL_SAMPLES
).
RESPONSE
2
RSP=1.11(0.00) SEC
Instrument response. Length of each signal processing
loop. Time in parenthesis is standard deviation.
VACUUM
5
VAC=9.1 IN-HG-A
Vacuum pressure.
SAMPPRESS
PRES=29.9 IN-HG-A
Sample pressure.
SAMPFLOW
SAMP FL=700 CC/M
Sample flow rate.
PMTDET
PMT=762.5 MV
Raw PMT reading.
NORMPMTDET
NORM PMT=742.9 MV
PMT reading normalized for temperature, pressure,
auto-zero offset, but not range.
UVDET
UV LAMP=3457.6 MV
UV lamp reading.
LAMPRATIO
LAMP RATIO=100.0 %
UV lamp ratio of current reading divided by calibrated
reading.
STRAYLIGHT
STR. LGT=0.1 PPB
Stray light offset.
DARKPMT
DRK PMT=19.6 MV
PMT dark offset.
DARKLAMP
DRK LMP=42.4 MV
UV lamp dark offset.
SLOPE
SLOPE=1.061
Slope for current range, computed during zero/span
calibration.
OFFSET
OFFSET=250.0 MV
Offset for current range, computed during zero/span
calibration.
HVPS
HVPS=650 VOLTS
High voltage power supply output.
RCELLDUTY
RCELL ON=0.00 SEC
Sample chamber temperature control duty cycle.
RCELLTEMP
RCELL TEMP=52.1 C
Sample chamber temperature.
BOXTEMP
BOX TEMP=35.5 C
Internal chassis temperature.
PMTTEMP
PMT TEMP=7.0 C
PMT temperature.
IZSDUTY
IZS ON=0.00 SEC
IZS temperature control duty cycle.
IZSTEMP
IZS TEMP=52.2 C
IZS temperature.
SO2 SO2=261.4
PPB
SO
2
concentration for current range.
TESTCHAN
TEST=3721.1 MV
Value output to
TEST_OUTPUT
analog output, selected
with
TEST_CHAN_ID
variable.
CLOCKTIME
TIME=10:38:27
Current instrument time of day clock.
1
The name is used to request a message via the RS-232 interface, as in “T BOXTEMP”.
2
Engineering
software.
3
Current instrument units.
5
M100EH
Summary of Contents for 100E
Page 2: ...Model 100E Instruction Manual ...
Page 4: ...Model 100E Instruction Manual ii 045150102 Rev XB1 ...
Page 12: ...Model 100E Instruction Manual x 045150102 Rev XB1 ...
Page 20: ...M100E Documentation Model 100E Instruction Manual 18 045150102 Rev XB1 User Notes ...
Page 26: ...Specifications Approvals and Warranty Model 100E Instruction Manual 24 045150102 Rev XB1 ...
Page 46: ...Getting Started Model 100E Instruction Manual 44 045150102 Rev XB1 User Notes ...
Page 128: ...Operating Instructions Model 100E Instruction Manual 126 045150102 Rev XB1 User Notes ...
Page 156: ...EPA Protocol Calibration Model 100E Instruction Manual 154 045150102 Rev XB1 ...