Theory Of Operation
Model 100E Instruction Manual
172
045150102 Rev XB1
10.3. Pneumatic Operation
Caution
It is important that the sample airflow system is leak-tight and not pressurized
over ambient pressure. Regular leak checks should be performed on the
analyzer as described in the maintenance schedule, Table 9-1 . Procedures for
correctly performing leak checks can be found in Section 11.5.1.
Relative Pressure versus Absolute Pressure
In this manual vacuum readings are given in inches of mercury absolute pressure (in-
Hg-A), i.e. indicate an absolute pressure referenced against zero (a perfect vacuum).
10.3.1.
Sample Gas Flow
The Flow of gas through the M100E UV Fluorescence SO2 Analyzer is created by a small
internal pump that pulls air though the instrument.
EXHAUST TO OUTER
LAYER OF KICKER
REA
C
T
IO
N
CEL
L
PU
RG
E
FLOW
SENSOR
FLOW / PRESSURE
SENSOR PCA
SAMPLE
FILTER
INSTRUMENT CHASSIS
EXHAUST GAS
OUTLET
VA
CU
UM
M
A
NI
F
O
L
D
SAMPLE GAS
INLET
KI
C
K
ER
EXHAU
S
T
T
O
PU
MP
PUMP
FLOW
CONTROL
ASSY
SAMPLE
PRESSURE
SENSOR
UV
LAMP
PMT
SAMPLE
CHAMBER
HYDROCARBON
SCRUBBER
(KICKER)
SPAN GAS INLET
ZERO AIR INLET
Figure 10-7: M100 Gas Flow and Location of Critical Flow Orifice
Summary of Contents for 100E
Page 2: ...Model 100E Instruction Manual ...
Page 4: ...Model 100E Instruction Manual ii 045150102 Rev XB1 ...
Page 12: ...Model 100E Instruction Manual x 045150102 Rev XB1 ...
Page 20: ...M100E Documentation Model 100E Instruction Manual 18 045150102 Rev XB1 User Notes ...
Page 26: ...Specifications Approvals and Warranty Model 100E Instruction Manual 24 045150102 Rev XB1 ...
Page 46: ...Getting Started Model 100E Instruction Manual 44 045150102 Rev XB1 User Notes ...
Page 128: ...Operating Instructions Model 100E Instruction Manual 126 045150102 Rev XB1 User Notes ...
Page 156: ...EPA Protocol Calibration Model 100E Instruction Manual 154 045150102 Rev XB1 ...