Getting Started
Model 100E Instruction Manual
42
045150102 Rev XB1
STEP THREE:
Perform the zero/span calibration procedure:
M-P CAL
STABIL=X.XXX PPB
SO2 =X.XXX
< TST TST >
ENTR
CONC
EXIT
M-P CAL
STABIL=X.XXX PPB
SO2 =X.XXX
< TST TST >
CAL
SETUP
M-P CAL
STABIL=X.XXX PPB
SO2 =X.XXX
< TST TST >
ZERO
CONC
EXIT
EXIT
returns to the main
SAMPLE
display
ACTION:
Allow zero gas to enter the sample port at the
rear of the instrument.
M-P CAL
STABIL=X.XXX PPB
SO2 =X.XXX
< TST TST >
SPAN
CONC
EXIT
M-P CAL RANGE = 500.0 PPB SO2 =X.XXX
< TST TST >
ENTR
SPAN CONC
EXIT
M-P CAL
RANGE = 500.0 PPB SO2 =X.XXX
< TST TST > ENTR CONC
EXIT
Wait until
STABIL
falls below
0.5 ppb.
This may take several
minutes.
The
SPAN
key now
appears during the
transition from zero to
span.
You may see both keys.
If either the
ZERO
or
SPAN
buttons fail to
appear see Section 11
for troubleshooting tips.
SAMPLE* RANGE = 500.000 PPB SO2 =X.XXX
< TST
TST
>
CAL
SETUP
SAMPLE
STABIL=X.XXX PPB
SO2 =X.XXX
< TST TST >
CAL
SETUP
Set the Display to show the
STABIL
test function.
This function calculates the
stability of the NO/NO
x
measurement
Press
ENTR
to changes the
OFFSET
&
SLOPE
values for the
SO
2
measurements.
Press
EXIT
to leave the calibration
unchanged and return to the
previous menu.
ACTION:
Allow span gas to enter the sample port at the
rear of the instrument.
The value of
STABIL
may jump
significantly.
Wait until it falls back
below
0.5 ppb.
This may take several
minutes.
Press
ENTR to
change the
OFFSET
&
SLOPE
values for the
SO
2
measurements.
Press
EXIT
to leave the calibration
unchanged and return to the
previous menu.
The Model 100E analyzer is now ready for operation.
Summary of Contents for 100E
Page 2: ...Model 100E Instruction Manual ...
Page 4: ...Model 100E Instruction Manual ii 045150102 Rev XB1 ...
Page 12: ...Model 100E Instruction Manual x 045150102 Rev XB1 ...
Page 20: ...M100E Documentation Model 100E Instruction Manual 18 045150102 Rev XB1 User Notes ...
Page 26: ...Specifications Approvals and Warranty Model 100E Instruction Manual 24 045150102 Rev XB1 ...
Page 46: ...Getting Started Model 100E Instruction Manual 44 045150102 Rev XB1 User Notes ...
Page 128: ...Operating Instructions Model 100E Instruction Manual 126 045150102 Rev XB1 User Notes ...
Page 156: ...EPA Protocol Calibration Model 100E Instruction Manual 154 045150102 Rev XB1 ...