Specifications, Approvals and Warranty
Model 100E Instruction Manual
20
045150102 Rev XB1
2.2. EPA Equivalency Designation
The Model 100E Analyzer is designated as Reference Method Number EQSA-0495-100 as
per 40 CFR Part 53 when operated under the following conditions:
•
Range: Any range from 50 parts per billion (ppb) to 10 parts per million (ppm).
•
Ambient temperature range of 5
o
C to 40
o
C.
•
Line voltage range of 105-125 VAC or 220-240 VAC, at 50 or 60 Hz.
•
Sample filter: Equipped with PTFE filter element in the internal filter assembly.
•
Sample flow of 650 +/- 65 cc/min.
•
Vacuum pump (internal or external) capable of 14"Hg Absolute pressure @ 1 slpm
or better.
•
Software settings:
Dynamic span
OFF
Dynamic zero
OFF
Dilution factor
OFF
AutoCal
ON or OFF
Dual range
ON or OFF
Auto-range
ON or OFF
Temp/Pressure compensation
ON
Under the designation, the analyzer may be operated with or without the following
optional equipment:
•
Rack mount with chassis slides.
•
Rack mount without slides, ears only.
•
Zero/span valve options.
•
Internal zero/span (IZS) option with:
a) SO
2
permeation tube - 0.4ppm at 0.7 liter per minute;
certified/uncertified.
b) SO
2
permeation tube - 0.8 ppm at 0.7 liter per minute;
certified/uncertified. Under the designation, the IZS option cannot be used
as the source of calibration.
•
4-20mA isolated analog outputs.
•
Status outputs.
•
Control inputs.
•
RS-232 output.
•
Ethernet output.
•
Zero air scrubber.
•
4-20mA, isolated output.
Summary of Contents for 100E
Page 2: ...Model 100E Instruction Manual ...
Page 4: ...Model 100E Instruction Manual ii 045150102 Rev XB1 ...
Page 12: ...Model 100E Instruction Manual x 045150102 Rev XB1 ...
Page 20: ...M100E Documentation Model 100E Instruction Manual 18 045150102 Rev XB1 User Notes ...
Page 26: ...Specifications Approvals and Warranty Model 100E Instruction Manual 24 045150102 Rev XB1 ...
Page 46: ...Getting Started Model 100E Instruction Manual 44 045150102 Rev XB1 User Notes ...
Page 128: ...Operating Instructions Model 100E Instruction Manual 126 045150102 Rev XB1 User Notes ...
Page 156: ...EPA Protocol Calibration Model 100E Instruction Manual 154 045150102 Rev XB1 ...