Calibration Procedures
Model 100E Instruction Manual
134
045150102 Rev XB1
Step Three:
Perform the calibration or calibration check according to the following flow
chart:
Wait until
STABIL
falls below 0.05
ppb. This may
take several
minutes.
ZERO CAL M
STABIL=XXX.X PPB
SO2 =X.XXX
< TST TST >
ZERO
CONC
EXIT
returns to the
main
SAMPLE
display
ZERO CAL M STABIL=XXX.X PPB
SO2 =X.XXX
< TST TST > ENTR CONC
EXIT
Pressing
ENTR
changes the calibration of the instrument.
If either the
ZERO
or
SPAN
button fails to
appear, see Chapter 11
for troubleshooting tips.
Analyzer enters
ZERO
CAL
mode
.
EXIT
returns the unit to
SAMPLE
mode without
changing the calibration
values.
ZERO CAL M STABIL=XXX.X PPB
SO2 =X.XXX
< TST TST > ZERO CONC
EXIT
ZERO CAL M STABIL=XXX.X PPB SO2=X.XX
X
< TST TST > CAL
CALZ
CALS
SETUP
SPAN CAL M
STABIL=XXX.X PPB
SO2 =X.XXX
< TST TST >
SPAN
CONC
EXIT
SPAN CAL M STABIL=XXX.X PPB
SO2 =X.XXX
< TST TST >
ENTR
CONC
EXIT
EXIT
returns to the
SAMPLE
mode without
changing the calibration
values.
SPAN CAL M STABIL=XXX.X PPB
SO2 =X.XXX
< TST TST >
SPAN
CONC
EXIT
ACTION:
Allow zero gas to enter the sample port at the
rear of the instrument.
SAMPLE* RANGE = 500.000 PPB SO2 =X.XXX
< TST
TST >
CAL
CALZ
CALS
SETUP
SAMPLE
STABIL=XXX.X PPB
SO2 =X.XXX
< TST TST > CAL
CALZ
CALS
SETUP
Scroll the display to the
STABIL
test function. This
function calculates the stability
of the NO/NO
X
measurements.
The value of
STABIL
may jump
significantly. Wait
until it falls below
0.5
ppb. This may take
several minutes.
Pressing
ENTR
changes the calibration of the instrument.
Analyzer enters
SPAN
CAL
Mode.
Summary of Contents for 100E
Page 2: ...Model 100E Instruction Manual ...
Page 4: ...Model 100E Instruction Manual ii 045150102 Rev XB1 ...
Page 12: ...Model 100E Instruction Manual x 045150102 Rev XB1 ...
Page 20: ...M100E Documentation Model 100E Instruction Manual 18 045150102 Rev XB1 User Notes ...
Page 26: ...Specifications Approvals and Warranty Model 100E Instruction Manual 24 045150102 Rev XB1 ...
Page 46: ...Getting Started Model 100E Instruction Manual 44 045150102 Rev XB1 User Notes ...
Page 128: ...Operating Instructions Model 100E Instruction Manual 126 045150102 Rev XB1 User Notes ...
Page 156: ...EPA Protocol Calibration Model 100E Instruction Manual 154 045150102 Rev XB1 ...