Model 100E Instruction Manual
Operating Instructions
045150102 Rev XB1
77
The following DC current output limits apply to the current loop modules:
Table 6-9: Analog Output Current Loop Range
Range
Minimum Output
Maximum Output
0-20 mA
0 mA
20 mA
These are the physical limits of the current loop modules, typical
applications use 2-20 or 4-20 mA for the lower and upper limits.
Please specify desired range when ordering this option.
The default offset for all ranges is 0 mA.
Pin assignments for the output connector at the rear panel of the instrument are shown
in Table 6-10.
ANALOG OUT
A1 A2 A3 A4
+ - + - + - + -
Table 6-10: Analog Output Pin Assignments
Pin Analog
output
VOLTAGE
Signal
CURRENT
Signal
1
V Out
I Out +
2
A1
Ground
I Out -
3
V Out
I Out +
4
A2
Ground
I Out -
5
A3
Not Used
Not Used
7
V Out
not available
8
A4
Ground not
available
See Figure 3-2 for a the location of the analog output connector on the instruments rear
panel.
Summary of Contents for 100E
Page 2: ...Model 100E Instruction Manual ...
Page 4: ...Model 100E Instruction Manual ii 045150102 Rev XB1 ...
Page 12: ...Model 100E Instruction Manual x 045150102 Rev XB1 ...
Page 20: ...M100E Documentation Model 100E Instruction Manual 18 045150102 Rev XB1 User Notes ...
Page 26: ...Specifications Approvals and Warranty Model 100E Instruction Manual 24 045150102 Rev XB1 ...
Page 46: ...Getting Started Model 100E Instruction Manual 44 045150102 Rev XB1 User Notes ...
Page 128: ...Operating Instructions Model 100E Instruction Manual 126 045150102 Rev XB1 User Notes ...
Page 156: ...EPA Protocol Calibration Model 100E Instruction Manual 154 045150102 Rev XB1 ...