Model 100E Instruction Manual
Calibration Procedures
045150102 Rev XB1
129
7.2. Manual Calibration
The following section describes the basic method for manually calibrating the Model 100E
SO
2
analyzer.
Calibration and Calibration Check
Pressing the
ENTR
key during the following procedure re-calculates the stored values
for OFFSET and SLOPE (instrument response curve) and alters the instrument’s
calibration.
If you wish to perform a calibration CHECK, do not press ENTR and see Section 7.3.
STEP ONE
: Connect the sources of zero air and span gas as shown below.
No Valve Options Installed
MODEL
100E
Sample
Zero Air
Exhaust
Span
MODEL 701
Zero Air
Generator
VENT
Calibrated
SO
2
GAS
(At high
concentration)
Source of
SAMPLE Gas
Removed
during
Calibration
Valve
OR
Needle valve to
control flow
MODEL
100E
Sample
Zero Air
Exhaust
Span
MODEL 701
Zero Air
Generator
VENT
Calibrated
SO
2
GAS
(At span gas
concentration)
Source of
SAMPLE Gas
Removed
during
calibration
MODEL 700 Gas
Dilution
Calibrator
(with Ozone Bench
Option)
Figure 7-1: Setup for Manual Calibration without Z/S valve or IZS Option
Summary of Contents for 100E
Page 2: ...Model 100E Instruction Manual ...
Page 4: ...Model 100E Instruction Manual ii 045150102 Rev XB1 ...
Page 12: ...Model 100E Instruction Manual x 045150102 Rev XB1 ...
Page 20: ...M100E Documentation Model 100E Instruction Manual 18 045150102 Rev XB1 User Notes ...
Page 26: ...Specifications Approvals and Warranty Model 100E Instruction Manual 24 045150102 Rev XB1 ...
Page 46: ...Getting Started Model 100E Instruction Manual 44 045150102 Rev XB1 User Notes ...
Page 128: ...Operating Instructions Model 100E Instruction Manual 126 045150102 Rev XB1 User Notes ...
Page 156: ...EPA Protocol Calibration Model 100E Instruction Manual 154 045150102 Rev XB1 ...