Model 100E Instruction Manual
Calibration Procedures
045150102 Rev XB1
133
7.4. Manual Calibration with Zero/Span Valves
Zero and Span calibrations using the Zero/Span Valve option are similar to that described
in Section 7.2, except that:
•
Zero air and span gas is supplied to the analyzer through the zero gas and span
gas inlets rather than through the sample inlet.
•
The zero and cal operations are initiated directly and independently with dedicated
keys (
CALZ
&
CALS
)
STEP ONE
: Connect the sources of zero air and span gas to the respective ports on the
rear panel (Figure 3-2) as shown below.
External Zero
Air Scrubber
Filter
MODEL 701
Zero Air
Generator
Source of
SAMPLE Gas
VENT if input is pressurized
MODEL 700
Gas Dilution Calibrator
(with O
3
generator option)
VENT
MODEL
100E
Sample
Zero Air
Exhaust
Span
Calibrated
SO
2
gas
(At high
concentration)
Figure 7-2: Setup for Manual Calibration with Z/S Valve Option Installed
Step Two:
Set the expected SO
2
span gas
value:
This sequence causes the
analyzer to prompt for the
expected NO and NO
x
span concentration.
M-P CAL
SO2 SPAN CONC: 450.0 Conc
0 0 0
4 5 .0
ENTR
EXIT
M-P CAL RANGE = 500.000 PPB SO2 =X.XXX
< TST TST > ZERO
CONC
EXIT
SAMPLE* RANGE = 500.000 PPB SO2 =X.XXX
< TST TST >
CAL
SETUP
The SO
2
span concentration
values automatically default
to
450.0 Conc
.
To change this value to the
actual concentration of the
span gas, enter the number
by pressing the key under
each digit until the expected
value appears.
The span gas concentration
should always be 90% of the
selected reporting range
EXAMPLE
Reporting range = 800 ppb
Span gas conc.= 720 ppb
EXIT
ignores the new setting
and returns to the previous
display.
ENTR
accepts the new setting
and returns to the
previous display..
Summary of Contents for 100E
Page 2: ...Model 100E Instruction Manual ...
Page 4: ...Model 100E Instruction Manual ii 045150102 Rev XB1 ...
Page 12: ...Model 100E Instruction Manual x 045150102 Rev XB1 ...
Page 20: ...M100E Documentation Model 100E Instruction Manual 18 045150102 Rev XB1 User Notes ...
Page 26: ...Specifications Approvals and Warranty Model 100E Instruction Manual 24 045150102 Rev XB1 ...
Page 46: ...Getting Started Model 100E Instruction Manual 44 045150102 Rev XB1 User Notes ...
Page 128: ...Operating Instructions Model 100E Instruction Manual 126 045150102 Rev XB1 User Notes ...
Page 156: ...EPA Protocol Calibration Model 100E Instruction Manual 154 045150102 Rev XB1 ...