Operating Instructions
Model 100E Instruction Manual
96
045150102 Rev XB1
6.9.6.
COM Port Communication Modes
Each of the analyzer’s serial ports can be configured to operate in a number of different
modes, which are listed in Table 6-17 and which can be combined by adding the Mode ID
numbers. For example, quiet mode, computer mode and internet-enabled mode would
carry a combined mode ID of 11, the standard configuration on the M100E COM2 port.
Note that each COM port needs to be configured independently.
Table 6-17: COMM Port Communication modes
MODE ID DESCRIPTION
QUIET MODE
1
Quiet mode suppresses any feedback from the analyzer (iDAS reports, and
warning messages) to the remote device and is typically used when the port
is communicating with a computer program such as APICOM. Such feedback
is still available but a command must be issued to receive them.
COMPUTER
MODE
2
Computer mode inhibits echoing of typed characters and is used when the
port is communicating with a computer program, such as APICOM.
SECURITY
MODE
4
When enabled, the serial port requires a password before it will respond.
The only command that is active is the help screen (? CR). See Section 6.4.1
for information on using the password feature.
ENABLE
INTERNET
8
Enables the use and configuration of the Ethernet interface. When enabled,
a new menu item INET will appear in the respective COM port menu.
Once this mode is enabled the
INET
sub menu will appear under the
COMMUNICATIONS MENU
(see Menu Tree A-5 in Appendix A-1)
HESSEN
PROTOCOL
16
The Hessen communications protocol is used in some European countries. T-
API part number 02252 contains more information on this protocol.
MULTIDROP
PROTOCOL
32
Multidrop protocol allows a multi-instrument configuration on a single
communications channel. Multidrop requires the use of instrument IDs.
ENABLE
MODEM
64
Enables to send a modem initialization string at power-up. Asserts certain
lines in the RS-232 port to enable the modem to communicate.
IGNORE
ERRORS
128
Fixes certain types of parity errors at certain Hessen protocol installations.
DISABLE
XON/XOFF
256
Disables XON/XOFF data flow control.
RS-485 MODE
1024
Configures the COM2 Port for RS-485 communication. RS-485 mode has
precedence over multidrop mode if both are enabled.
COMMAND
PROMPT
4096
Enables a command prompt when in terminal mode.
Summary of Contents for 100E
Page 2: ...Model 100E Instruction Manual ...
Page 4: ...Model 100E Instruction Manual ii 045150102 Rev XB1 ...
Page 12: ...Model 100E Instruction Manual x 045150102 Rev XB1 ...
Page 20: ...M100E Documentation Model 100E Instruction Manual 18 045150102 Rev XB1 User Notes ...
Page 26: ...Specifications Approvals and Warranty Model 100E Instruction Manual 24 045150102 Rev XB1 ...
Page 46: ...Getting Started Model 100E Instruction Manual 44 045150102 Rev XB1 User Notes ...
Page 128: ...Operating Instructions Model 100E Instruction Manual 126 045150102 Rev XB1 User Notes ...
Page 156: ...EPA Protocol Calibration Model 100E Instruction Manual 154 045150102 Rev XB1 ...