Performance Veri
fi
cation
Push menu button
Select menu option
Select setting
Trig Menu
Slope
Falling
15.
Push
Set To 50%
. Adjust
Trigger Level
as necessary and then check that
triggering is stable.
16.
Disconnect the test setup.
17.
Repeat steps 1 through 16 for all input channels.
Check External Edge
Trigger Sensitivity
This test checks the edge trigger sensitivity for the external trigger.
1.
Set up the oscilloscope using the following table:
Push menu button
Select menu option
Select setting
Default Setup
—
—
Channel
1
Probe
1X
Source
Ext
Trig Menu
Mode
Normal
Acquire
Sample
—
Source
CH1
Measure
Type
Pk-Pk
2.
Connect the oscilloscope to the leveled sine wave generator as shown in the
following
fi
gure, using channel
1
and
Ext Trig
.
3.
Set the oscilloscope
Vertical Scale
(volts/division) to
100 mV/div
.
4.
Set the oscilloscope
Horizontal Scale
(seconds/division) to
25 ns/div
.
5.
Set the leveled sine wave generator frequency to
10 MHz
.
6.
Set the sine wave generator output level to approximately
300 mV
p-p
into the
power splitter. This is about
200 mV
p-p
on channel
1
of the oscilloscope.
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
35
Summary of Contents for TDS1001C-EDU
Page 12: ...Preface x TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...
Page 30: ...Specifications 18 TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...
Page 52: ...Performance Verification 40 TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...
Page 60: ...Adjustment Procedures 48 TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...
Page 85: ...Maintenance TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual 73...
Page 96: ...Maintenance 84 TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...
Page 110: ...Replaceable Parts 98 TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...