Adjustment Procedures
Adjustment Procedure
Use this procedure to load new calibration constants.
1.
Enable the Service menu. (See page 42,
2.
Push the
Service
option button.
3.
Push the
Factory Cal
option button.
4.
Push the
Warmup Timer
option button.
5.
Push the
Start Timer
option button to allow a 10-minute warm up. After
10 minutes, the oscilloscope displays a large black or green cross.
6.
Press the front-panel
Utility
button and push the
Service
option button.
7.
Push the
Factory Cal
option button.
8.
Push the
FCAL INIT
option button to initiate the factory calibration routine.
You are now in step 1 of the procedure.
9.
As indicated by step 1, connect a 0.0 V
DC
source to the channel inputs. (See
10.
Press
FCAL STEP
to load the calibration constant for step 1.
11.
Follow the screen prompts and the Adjustment setup table for the remainder
of the routine, applying the proper signal for each step. (See Table 15.)
NOTE.
During some steps, the oscilloscope may appear to be idle for several
minutes while it is processing information internally. You must wait for the screen
prompt to continue.
If any step in the procedure fails, the procedure terminates. Failure will result
if an incorrect signal source is connected. If this happens, you must start the
procedure over from the beginning.
If the procedure completes successfully, a "Pass" message is displayed and the
new adjustment takes effect.
12.
Compensate the signal path by performing the self-calibration routine.
13.
Disconnect all probes and cables from the channel input connectors.
14.
Press the
Utility
button and select
Do Self Cal
to con
fi
rm that you are ready
to proceed.
15.
Disable the Service menu option by pushing the
Utility
front-panel button and
the
Service
and
Service Mode Off
option buttons.
44
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Summary of Contents for TDS1001C-EDU
Page 12: ...Preface x TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...
Page 30: ...Specifications 18 TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...
Page 52: ...Performance Verification 40 TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...
Page 60: ...Adjustment Procedures 48 TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...
Page 85: ...Maintenance TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual 73...
Page 96: ...Maintenance 84 TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...
Page 110: ...Replaceable Parts 98 TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...