Speci
fi
cations
Table 2: Time base system (cont.)
Characteristic
Description
This is the accuracy of delta time measurements made on any single waveform.
The limits are given in the following table for signals having an amplitude
≥
5 divisions, a slew rate at the
measurement points of
≥
2.0 divisions/ns, and acquired
≥
10 mV/div.
Condition
Time Measurement Accuracy
Single shot, sample mode,
full bandwidth selected
±(1 Sample In 100 ppm * |reading| + 0.6 ns)
> 16 averages,
full bandwidth selected
±(1 Sample In 100 ppm * |reading| + 0.4 ns)
Delta Time Measurement
Accuracy
The Sample Interval is the time between the samples in the waveform record.
Time Measurement
Functions
Frequency, Period, Rise Time, Fall Time, Positive Width, negative Width, Positive Duty Cycle, Phase,
Delay
8
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Summary of Contents for TDS1001C-EDU
Page 12: ...Preface x TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...
Page 30: ...Specifications 18 TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...
Page 52: ...Performance Verification 40 TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...
Page 60: ...Adjustment Procedures 48 TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...
Page 85: ...Maintenance TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual 73...
Page 96: ...Maintenance 84 TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...
Page 110: ...Replaceable Parts 98 TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...