Maintenance
Keypad
You will need a torque-limiting Torx T-15 screwdriver for this procedure.
Refer to the exploded view diagram for the 2-channel models or the 4-channel
models. (See page 89.)(See Figure 13.)
Removal.
1.
Remove the power button and rear case. (See page 58,
2.
Remove the internal assembly from the front case. (See page 61,
3.
Lift the keypad switch out of the front case.
CAUTION.
Keep the electrical contacts on the back of the keypad clean. Skin oils
and dust inhibit good electrical contact.
Installation.
1.
Place the front case face down on a soft surface (such as an anti-static mat).
2.
Align the switch keypad into the front case using the guide posts and the
holes in the keypad.
CAUTION.
Keep the electrical contacts on the back of the switch keypad clean.
Skin oils and dust inhibit good electrical contact.
3.
Check that all buttons are properly seated before proceeding.
4.
Assemble the oscilloscope by performing the installation procedure in reverse
order for each module that was removed.
68
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Summary of Contents for TDS1001C-EDU
Page 12: ...Preface x TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...
Page 30: ...Specifications 18 TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...
Page 52: ...Performance Verification 40 TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...
Page 60: ...Adjustment Procedures 48 TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...
Page 85: ...Maintenance TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual 73...
Page 96: ...Maintenance 84 TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...
Page 110: ...Replaceable Parts 98 TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...