Maintenance
Support the BNC to prevent it from twisting the Main board, and torque the
nuts to 40 in-lbs (4.5 N@m) using the torque wrench.
3.
Install the three screws attaching the Main board module to the internal
assembly.
4.
Connect the following cables on the Main board:
The power supply cable at J101
The display ribbon cable (J1501 for 2 Ch, J2002 for 4 Ch)
The backlight wire cable (J1502 for 2 Ch, J2001 for 4 Ch)
The front-panel cable at J202
5.
Assemble the oscilloscope by performing the installation procedure in reverse
order for each module that was removed.
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual