Performance Veri
fi
cation
3.
Connect the oscilloscope channel under test to the DC voltage source as
shown in the following
fi
gure:
4.
For each vertical scale (volts/division) setting in the following table, perform
the following steps:
a.
Set the DC voltage source output level to the positive voltage listed and
then record the mean measurement as V
pos
.
b.
Reverse the polarity of the DC voltage source and record the mean
measurement as V
neg
.
c.
Calculate V
diff
= V
pos
- V
neg
and compare V
diff
to the accuracy limits in
the following table:
Vertical Scale
(volts/div) setting
DC voltage
source output levels
Accuracy limits for V
diff
5 mV/div
+17.5 mV, -17.5 mV
33.6 mV to 36.4 mV
200 mV/div
+700 mV, -700 mV
1.358 V to 1.442 V
2 V/div
+7.00 V, -7.00 V
13.58 V to 14.42 V
5.
Set DC voltage source output level to
0 V
.
6.
Disconnect the test setup.
7.
Repeat steps 1 through 3 for all input channels.
30
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Summary of Contents for TDS1001C-EDU
Page 12: ...Preface x TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...
Page 30: ...Specifications 18 TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...
Page 52: ...Performance Verification 40 TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...
Page 60: ...Adjustment Procedures 48 TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...
Page 85: ...Maintenance TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual 73...
Page 96: ...Maintenance 84 TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...