Speci
fi
cations
Table 3: Triggering system (cont.)
Characteristic
Description
Trigger Source
Sensitivity
AC
Same as DC Coupled limits for frequencies 50 Hz and
above
Noise Rej
Effective in Sample or Average Mode, > 10 mV/div to
5 V/div. Reduces DC Coupled trigger sensitivity by 2X.
HF Rej
Same as DC Coupled limits from DC to 7 kHz.
LF Ref
Same as DC Coupled limits for frequencies above 300 kHz.
Sensitivity,
Edge-Type Trigger,
non-DC Coupled,
typical
Since AC coupling is not done in the front end, use of a 10M probe does not affect the low frequency
corner.
50 Hz.
Lowest Frequency
for Successful
Operation of “Set
Level to 50%”
Function, typical
Using a 10M probe will not affect the operation of this function.
Input Channel
±8 divisions from center screen
Ext
±1.6 V
Ext/5:
±8 V
Trigger Level
Ranges, typical
The settable resolution for Trigger Level is 0.02 division for an input channel source, 4 mV for Ext source,
and 20 mV for Ext/5 source.
This is the amount of deviation allowed between the level on the waveform at which triggering occurs and
the level selected for DC-coupled triggering signals
±(0.2 div + 5 mV) for signals within ±4 divisions from the center screen, having rise and fall times
of
≥
20 ns.
Ext: ±(6% of s 40 mV) for signals less than ±800 mV
Trigger Level
Accuracy, DC
Coupled, typical
Ext/5: ±(6% of s 200 mV) for signals less than ±4 V
Video Trigger
Trigger Mode: Auto
Default Settings
for Video Trigger
Trigger Coupling: AC
Video Trigger
Source Selection
Same as Trigger Source Selections listed above except Line Trigger. Line Trigger source is meaningless
in this mode.
Video Trigger
Polarity Selection
Normal (Negative going Sync Signal), Invert (Positive going Sync Signal)
Video Sync
Selection
Line, Line #, Odd Field, Even Field, Field: PAL/SECAM, NTSC formats
Field rates: 50 Hz to 60 Hz.
Video Trigger
Formats and Field
Rates
Line rates: 15 kHz to 20 kHz (NTSC, PAL, SECAM)
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
11
Summary of Contents for TDS1001C-EDU
Page 12: ...Preface x TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...
Page 30: ...Specifications 18 TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...
Page 52: ...Performance Verification 40 TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...
Page 60: ...Adjustment Procedures 48 TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...
Page 85: ...Maintenance TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual 73...
Page 96: ...Maintenance 84 TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...
Page 110: ...Replaceable Parts 98 TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...