Maintenance
Troubleshooting the
Display
To troubleshoot a nonfunctional display, follow these steps:
1.
Remove the rear case. (See page 58,
2.
Remove the internal assembly from the front case. (See page 61,
3.
Remove the Display module. (See page 66,
4.
Reseat the display cable to the Display module, making sure that the cable is
fully seated in the connector.
5.
Reconnect the Display module to the internal assembly.
6.
Reattach the front panel.
7.
Reconnect the backlight and display cables to the Main board.
8.
Power on the oscilloscope. If the display still does not work, contact your
local Tektronix Customer Service Center for assistance.
76
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual