Performance Veri
fi
cation
10.
Check that the rising edge of the marker crosses the center horizontal graticule
line within ±2 divisions of the vertical center graticule line, as shown in the
following
fi
gure:
NOTE.
One division of displacement from graticule center corresponds to a
25 ppm time base error.
11.
Disconnect the test setup.
Check Edge Trigger
Sensitivity
This test checks the edge trigger sensitivity for all input channels.
1.
Set up the oscilloscope using the following table:
Push menu button
Select menu option
Select setting
Default Setup
—
—
Channel
1
Probe
1X
Trig Menu
Mode
Normal
Acquire
Sample
—
Source
Channel under test
Measure
Type
Pk-Pk
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
33