Operation
Mask testing
The analyzer lets you test many trace results for changes in level. You can identify matches crossing a power level or you
can create a mask that speci
fi
es power level and frequency range limits. You can select which trace to test and specify
an action to take when a test condition is met for traces in the Spectrum, DPX, Spurious, and Setting Time displays.
The Frequency Settling Time, Phase Settling Time, Spurious, Bluetooth InBand Emission, and Occupied Bandwidth
measurements tests provide pass/fail results.
The available choices include only results from displays that are currently open.
Selecting
Mask Test
from the Tools menu displays the Mask Test control panel. These settings de
fi
ne test parameters and
specify actions to be performed when the test conditions are met.
To start a mask test.
Perform the following steps to set up for any mask test.
1.
Open the Spectrum display or select the
Spectrum display if it is already open.
2.
Select
Tools
>
Mask Search
.
3.
Select the
Enable Test
check box to set
the application to perform the test once
the conditions are de
fi
ned.
4.
Select the trace you wish to test from the
Search in
drop-down list box.
RSA7100 Series Quick Start User Manual