Operation
Measurements with con
fl
icting acquisition requirements
The analyzer lets you open any combination of measurements. You can select the displays that you are interested in and the
analyzer will automatically set up the appropriate analysis and acquisition parameters.
If measurements place con
fl
icting requirements on the acquisition parameters, the current selected display takes priority. The
following example shows incompatible measurement settings when one display needs real-time data (such as the Amplitude
vs Time measurement) while another display requires swept data (such as a Spectrum graph with a very wide span).
1.
Click the
Preset
button (Main preset) to
set the analyzer to the default settings.
2.
Tune the analyzer to the frequency of
your signal and then set the span.
3.
Click the Displays icon to open the Select
Displays dialog box.
4.
Select the
General Signal Viewing
folder.
5.
Select the
Amplitude vs Time
icon.
6.
Click
Add
to add the icon to the Selected
Displays list.
NOTE.
Add the Spectrum display if it is not
already in the Selected Displays list.
7.
Click
OK
to accept the changes and
close the dialog box.