![Tektronix RSA3300A Service Manual Download Page 66](http://html1.mh-extra.com/html/tektronix/rsa3300a/rsa3300a_service-manual_1078139066.webp)
Performance Tests
4- 10
RSA3303A & RSA3308A Service Manual
Prerequisites
The tests in this section comprise an extensive, valid confirmation of perfor-
mance and functionality when the following requirements are met:
H
The cabinet must be installed on the analyzer.
H
The instrument must have passed the procedures under
Brief Procedures
found on page 4--3.
H
A signal-path compensation must have been done within the recommended
calibration interval and at a temperature within
±
3
_
C of the present
operating temperature. (If at the time you did the prerequisite
Brief Proce-
dures,
the temperature was within the limits just stated, consider this
prerequisite met.)
H
The analyzer must have been last adjusted at an ambient temperature
b20
_
C and +30
_
C, must have been operating for a warm-up
period of at least 20 minutes, and must be operating at an ambient tempera-
ture of b10
_
C and +40
_
C.
H
The SG flatness/spurious floppy disk(s) (Item 17 in Table 4--3) must be
prepared. For the procedure to create or update the flatness and spurious
correction files, refer to
Making Flatness Correction Files
on page 5--6 and
Making Spurious Correction Files
on page 5--13.
Equipment Required
These procedures use external, traceable signal sources to directly check
warranted characteristics. Table 4--3 shows the required equipment.
Table 4- 3: Test equipment
Item number and
description
Minimum requirements
Example
Purpose
1. Signal generator
1
Frequency range : 10 kHz to 3.5 GHz;
Variable amplitude from --70 dBm to
+13 dBm into 50
Ω
;
Accuracy: <
±
1 dB; Function: FM
Agilent 8648D
Checking flatness in base-
band and RF1 band
2. Signal generator
Frequency range: 10 MHz to 8 GHz;
Output level: --30 dBm to +10 dBm;
Accuracy: <
±
1 dB
Agilent 83712B with
option 1E1
Checking flatness in RF2 and
RF3 bands (RSA3308A only)
3. Signal generator
Frequency range: 1 MHz to 8 GHz;
Output level: --30 dBm to +10 dBm;
Accuracy: <
±
1 dB
Agilent E8257D with
option 1E1
(option UNX recommended)
Checking flatness in all bands
1
For RSA3303A, you can use the signal generator with the frequency range of 10 kHz to 3 GHz, for example Agilent 8648C.
Summary of Contents for RSA3300A
Page 4: ......
Page 12: ...Table of Contents viii RSA3303A RSA3308A Service Manual...
Page 16: ...Service Safety Summary xii RSA3303A RSA3308A Service Manual...
Page 20: ...Preface xvi RSA3303A RSA3308A Service Manual...
Page 22: ...Introduction xviii RSA3303A RSA3308A Service Manual...
Page 23: ...Specifications...
Page 24: ......
Page 43: ...Operating Information...
Page 44: ......
Page 46: ...Operating Information 2 2 RSA3303A RSA3308A Service Manual...
Page 47: ...Theory of Operation...
Page 48: ......
Page 55: ...Performance Verification...
Page 56: ......
Page 109: ...Adjustment Procedures...
Page 110: ......
Page 145: ...Maintenance...
Page 146: ......
Page 214: ...Removal and Installation Procedures 6 68 RSA3303A RSA3308A Service Manual...
Page 231: ...Options...
Page 232: ......
Page 234: ...Options and Accessories 7 2 RSA3303A RSA3308A Service Manual...
Page 235: ...Electrical Parts List...
Page 236: ......
Page 238: ...Electrical Parts List 8 2 RSA3303A RSA3308A Service Manual...
Page 239: ...Diagrams...
Page 240: ......
Page 242: ...Diagrams 9 2 RSA3303A RSA3308A Service Manual...
Page 249: ...Mechanical Parts List...
Page 250: ......
Page 283: ......
Page 284: ......