Adjustment Procedures
5- 30
RSA3303A & RSA3308A Service Manual
2.
Adjust the reference frequency:
Adjust R232 on the A500 step amplifier/
oscillator interface board in the RF5 module so that the frequency counter
reads 10 MHz
0.1 Hz. See Figure 5--13 for the R232 location.
RF5 module
R232
Rear
W32
W29
Figure 5- 13: R232 location (top view of the card cage)
3.
Restore the analyzer to normal operation:
a.
Power off the analyzer.
b.
Refer to
Removal and Installation Procedures
in Section 6 to reinstall
the cabinet and other modules removed in step 1.
Summary of Contents for RSA3300A
Page 4: ......
Page 12: ...Table of Contents viii RSA3303A RSA3308A Service Manual...
Page 16: ...Service Safety Summary xii RSA3303A RSA3308A Service Manual...
Page 20: ...Preface xvi RSA3303A RSA3308A Service Manual...
Page 22: ...Introduction xviii RSA3303A RSA3308A Service Manual...
Page 23: ...Specifications...
Page 24: ......
Page 43: ...Operating Information...
Page 44: ......
Page 46: ...Operating Information 2 2 RSA3303A RSA3308A Service Manual...
Page 47: ...Theory of Operation...
Page 48: ......
Page 55: ...Performance Verification...
Page 56: ......
Page 109: ...Adjustment Procedures...
Page 110: ......
Page 145: ...Maintenance...
Page 146: ......
Page 214: ...Removal and Installation Procedures 6 68 RSA3303A RSA3308A Service Manual...
Page 231: ...Options...
Page 232: ......
Page 234: ...Options and Accessories 7 2 RSA3303A RSA3308A Service Manual...
Page 235: ...Electrical Parts List...
Page 236: ......
Page 238: ...Electrical Parts List 8 2 RSA3303A RSA3308A Service Manual...
Page 239: ...Diagrams...
Page 240: ......
Page 242: ...Diagrams 9 2 RSA3303A RSA3308A Service Manual...
Page 249: ...Mechanical Parts List...
Page 250: ......
Page 283: ......
Page 284: ......