Perform ance Check Procedure—2246 1Y and 2246 Mod A Service
h. Switch CH 2 VOLTS/DIV through all positions
from 2 mV to 5 V.
i.
CHECK—trace shift does not exceed 0.2 division
between steps.
3. CH 3 and CH 4 VOLTS/DIV Trace Shift
a. Set Vertical MODE to CH 3 (CH 2 off).
b. Position trace to the center horizontal graticule
line.
c. Switch CH 3 VOLTS/DIV between 0 1 V and
0.5 V.
d. CHECK—trace
shift
does not exceed
one
division.
e. Set Vertical MODE to CH 4 (CH 3 off).
f.
Position trace to the center horizontal graticule
line,
g. Switch CH 4 VOLTS/DIV between 0.1 V and
0.5 V.
h. CHECK—trace
shift
does not exceed one
division.
4. CH 1 and CH 2 VAR VOLTS/DIV Trace Shift
a. Set:
Vertical MODE
CH 1 (CH 4 off)
CH 1 VOLTS/DIV
2 mV
b. Position trace to center graticule line.
c. Set CH 1 VAR VOLTS/DIV fully CCW.
d. CHECK—trace shift does not exceed one
division.
e. Set:
CH 1 VAR VOLTS/DIV
Detent
(calibrated)
Vertical MODE
CH 2 (CH 1 off)
CH 2 VOLTS/DIV
2 mV
f.
Position trace to center graticule line.
g. Set CH 2 VAR VOLTS/DIV fully CCW.
h. CHECK-trace shift does not exceed one
division.
i.
Set CH 2 VAR VOLTS/DIV to detent (calibrated)
position.
5. CH 1 and CH 2 Input COUPLING Trace Shift
a. Position trace to center graticule line.
b. Set CH 2 Input COUPLING to DC.
c. CHECK—trace shift does not exceed 0.25
division.
d. Set:
Vertical MODE
CH 1 (CH 2
off)
CH 1 Input COUPLING
GND
e. Position trace to center graticule line.
f.
Set CH 1 Input COUPLING to DC.
g. CHECK-trace shift does not exceed 0.25
division.
6. CH 2 INVERT Trace Shift
a. Set:
Vertical MODE to CH 2
(CH 1 off)
CH 2 Input Coupling
GND
b. Position trace to center horizontal graticule line.
c. Set CH 2 INVERT On.
d. CHECK-trace shift does not exceed one
division.
e. Set:
CH 2 INVERT
Off
CH 2 COUPLING
DC
7. CH 1 and CH 2 VAR VOLTS/DIV Range
a. Set Vertical MODE to CH 1.
b. Position CH 1 trace to the center horizontal
graticule line,
c. Connect calibration generator (PG 506) STD
AMPL output to the CH 1 input via 50 G precision
BNC coaxial cable. Set calibration generator to
STD AMPL and AMPLITUDE to 50 mV.
d. Set:
CH 1 VOLTS/DIV
10
mV
CH 1 VAR VOLTS/DIV
Fully CCW
e. CHECK—the signal amplitude is two divisions or
less.
4-7
Summary of Contents for 2246 1Y
Page 13: ...2246A Service 7062 01 X The 2246 1Y or 2246 Mod A Portable Oscilloscope ...
Page 35: ......
Page 139: ......
Page 185: ......
Page 187: ......
Page 278: ...Figure 9 5b Detailed 2246 1Y or 2246 Mod A block diagram part 2 R E ...
Page 287: ......
Page 304: ...A D D J U N 1991 ...
Page 311: ...2246 1Y and 2246 Mod A Service WAVEFORMS FOR DIAGRAM 1 0 V r l III B f e I l i 6555 39 ...
Page 319: ......
Page 320: ......
Page 321: ...2246 1Y and 2246 Mod A Service WAVEFORMS FOR DIAGRAM 2 SET READOUT CONTROL CCW OFF 6555 40 ...
Page 323: ......
Page 324: ......
Page 326: ...A B C P E F G ...
Page 327: ......
Page 331: ......
Page 337: ......
Page 340: ......
Page 343: ...C l rti ...
Page 347: ...BEV JUNE 1991 7062 18 ...
Page 349: ...2246 1Y and 2246 Mod A Service WAVEFORMS FOR DIAGRAM 6 6555 43 I ...
Page 355: ...I BOARD LOCATION 8K 8J 8H 7H 8H 8H 7J 8D 7H 8H 8J 8J 7K 8K 10D 6L 6K 6L 6K 6G 5H I ...
Page 362: ......
Page 372: ...2246 1Y and 2246 Mod A Service WAVEFORMS FOR DIAGRAM 9 MORE ...
Page 373: ... 6555 72 6557 88 ...
Page 375: ......
Page 380: ...WAVEFORMS FOR DIAGRAM 10 ...
Page 387: ...WAVEFORMS FOR DIAGRAM 11 6555 73 6557 48 ...
Page 394: ...WAVEFORMS FOR DIAGRAM 12 2246 1Y and 2246 Mod A Service i f I S 1 f c i 4 i ...
Page 395: ...WAVEFORMS FOR DIAGRAM 12 cont 2246 1Y and 2246 Mod A Service ...
Page 397: ......
Page 398: ......
Page 400: ...REV JUNE 1991 7062 39 ...
Page 401: ......
Page 403: ......
Page 404: ......
Page 406: ...B 2246 IV AND 2246 MOD A ...
Page 407: ......
Page 408: ......
Page 414: ... _ 5 ...
Page 415: ...2246 1Y 2246 MOD A ...
Page 416: ......
Page 423: ...2246 1Y 2246 MOD A ...
Page 424: ...2246 1Y 2246 MOD A ...
Page 425: ......
Page 433: ......