M aintenance—2246 1Y and 2246 Mod A Service
2. ROM diagnostics—failures indicated by flashing
AUTO lamp. Tests all of ROM, except for ROM
header. Runs proprietary version of CROC test.
3. Readout interface diagnostics—failures indicated
by flashing NORM lamp. Performs a m arching-
one test around the loop from D7-D0 to R7-R0
and back.
SERVICE MODE
The service mode driver menu is accessed by
pressing the top and the bottom m enu-item buttons
at the same time. The initial SERVICE MENU will be
displayed as shown in Figure 6-2. Each service
menu display has two parts; the part to the left is the
service menu, and the part to the right is the
modifier menu.
the left side of the screen with the name of the
selected sub-m enu displayed in the top line.
A menu choice that has no following slash is an
executable service routine. The routine may be run
by underlining it and pressing the menu button next
to the RUN label that appears in the modifier menu
list. Executable servicing selections are: diagnostics
that return either a pass message or a fail message
along with service data; one-shot exercisers that
carry out some service and immediately return to
the menu; or regular exercisers that carry out a
service while continuously displaying service data.
An underlined service-m enu choice is available for
selection. To select a menu item, use the buttons
next to the up-arrow and down-arrow symbols to
move the underline up or down in the SERVICE
MENU. When the underline is below the sub-menu
title, pressing the menu button next to up-arrow
MENU label returns to the preceding menu con
taining that sub-m enu (an up-m enu operation).
SERVICE MENU/
DIAGNOSE
t MENU
CONFIGURE
4,
SELF CAL MEASUREMENTS
INTERNAL SETTINGS MENU/
EXERCISER MENU/
QUIT
6555-23
Figure 6-2. Initial SERVICE MENU.
Each service menu has a title and a number of
selectable items in the menu. The title appears in
the top line of the crt display, and the selectable
items appear under the title,
indented
(see
Figure 6 -2 ). Menu choices that are names of sub
menus have a following slash (/), and when one is
underlined, the word “ SELECT” appears in the
modifier menu list. When a choice with sub-menus
i s s e l e c t e d , t h e s u b - m e n u c h o i c e s
are displayed on
Pressing the QUIT selection or CLEAR Display button
at any time the choice is displayed will cause the
scope to return to normal oscilloscope mode. If a
service routine is operating that has an END menu
selection displayed, pressing the button next to END
exits the routine and returns to the selection menu
(where QUIT is displayed). Routines that run once,
return to the selection menu when finished.
The diagnostic tests in the SERVICE MENU may be
run with a conditional setting that determines how
many times the routine is done. The conditional
MODE setting menu choice appears in the modifier
menu when the DIAGNOSE choice in the SERVICE
MENU is underlined (see Figure 6 -3 ). One of the fol
lowing mode types will be displayed:
ONCE, CONTINUOUS, UNTIL PASS, or UNTIL FAIL
Change the mode type displayed in the bottom line
by pressing the MODE button. When ONCE is the
mode, the diagnostic is run once, and the result is
displayed. When CONTINUOUS is the mode, the
diagnostics are run continuously. When UNTIL PASS
is the mode, the diagnostics are run until they pass.
When UNTIL FAIL is the mode, the diagnostics are
run until they fail. In order to stop a diagnostic that is
looping in the CONTINUOUS, UNTIL PASS, or UNTIL
FAIL mode, press the HALT button. The diagnostic
will stop and display the current status. When the
status is displayed, press END to return to the
SERVICE MENU choices.
6-11
Summary of Contents for 2246 1Y
Page 13: ...2246A Service 7062 01 X The 2246 1Y or 2246 Mod A Portable Oscilloscope ...
Page 35: ......
Page 139: ......
Page 185: ......
Page 187: ......
Page 278: ...Figure 9 5b Detailed 2246 1Y or 2246 Mod A block diagram part 2 R E ...
Page 287: ......
Page 304: ...A D D J U N 1991 ...
Page 311: ...2246 1Y and 2246 Mod A Service WAVEFORMS FOR DIAGRAM 1 0 V r l III B f e I l i 6555 39 ...
Page 319: ......
Page 320: ......
Page 321: ...2246 1Y and 2246 Mod A Service WAVEFORMS FOR DIAGRAM 2 SET READOUT CONTROL CCW OFF 6555 40 ...
Page 323: ......
Page 324: ......
Page 326: ...A B C P E F G ...
Page 327: ......
Page 331: ......
Page 337: ......
Page 340: ......
Page 343: ...C l rti ...
Page 347: ...BEV JUNE 1991 7062 18 ...
Page 349: ...2246 1Y and 2246 Mod A Service WAVEFORMS FOR DIAGRAM 6 6555 43 I ...
Page 355: ...I BOARD LOCATION 8K 8J 8H 7H 8H 8H 7J 8D 7H 8H 8J 8J 7K 8K 10D 6L 6K 6L 6K 6G 5H I ...
Page 362: ......
Page 372: ...2246 1Y and 2246 Mod A Service WAVEFORMS FOR DIAGRAM 9 MORE ...
Page 373: ... 6555 72 6557 88 ...
Page 375: ......
Page 380: ...WAVEFORMS FOR DIAGRAM 10 ...
Page 387: ...WAVEFORMS FOR DIAGRAM 11 6555 73 6557 48 ...
Page 394: ...WAVEFORMS FOR DIAGRAM 12 2246 1Y and 2246 Mod A Service i f I S 1 f c i 4 i ...
Page 395: ...WAVEFORMS FOR DIAGRAM 12 cont 2246 1Y and 2246 Mod A Service ...
Page 397: ......
Page 398: ......
Page 400: ...REV JUNE 1991 7062 39 ...
Page 401: ......
Page 403: ......
Page 404: ......
Page 406: ...B 2246 IV AND 2246 MOD A ...
Page 407: ......
Page 408: ......
Page 414: ... _ 5 ...
Page 415: ...2246 1Y 2246 MOD A ...
Page 416: ......
Page 423: ...2246 1Y 2246 MOD A ...
Page 424: ...2246 1Y 2246 MOD A ...
Page 425: ......
Page 433: ......