Performance Check Procedure —2246 1Y and 2246 Mod A Service
f.
Set:
CH 1 VAR VOLTS/DIV
Detent
(calibrated)
Vertical MODE
CH 2 (CH 1 off)
CH 2 VOLTS/DIV
10 mV
g. Position CH 2 trace to the center horizontal graticule
line.
h. Move the test signal to the CH 2 input.
f.
Move the test signal to the CH 1 input.
g. Set:
Vertical MODE
CH 1 (CH 2 off)
CH 1 VOLTS/DIV
2 mV
h. Position the trace two graticule lines below the cen
ter graticule line.
i. Repeat CHECK procedure for CH 1.
i.
Set CH 2 VAR VOLTS/DIV
fully CCW.
j.
Repeat the CHECK procedure for CH 2.
k. Set CH 2 VAR VOLTS/DIV to detent (calibrated) posi
tion.
8. Low-Frequency Linearity
a. Set:
Vertical MODE
CH 1 (CH 2 off)
CH 1 VOLTS/DIV
10 mV
SCOPE BW
On
b. Set calibration generator AMPLITUDE to 20 mV.
c. Move the test signal to the CH 1 input.
d. Position the top of the signal to top graticule line.
e. Check the signal amplitude is between 1.9 and 2.1
divisions.
f.
Set bottom of the signal to bottom graticule line.
g. Check the signal amplitude is between 1.9 and 2.1
divisions.
h.
Repeat the procedure for CH 2.
9. CH 1 and CH 2 Vertical Deflection Accuracy
a. Set CH 2 VOLTS/DIV to 2 mV.
b. Set calibration generator AMPLITUDE to 10 mV.
c.
Position the trace two graticule lines below the cen
ter graticule line.
d. CHECK—all positions of the VOLTS/DIV settings for
correct signal-to-graticule accuracy, using the set
tings in Table 4-2 for the checks.
e. Set calibration generator to Std Ampl output, 10 mV.
Table 4 -2
S ignal-to-G raticule Accuracy
VOLTS/DIV
Setting
Calibration
Generator
AMPLITUDE Setting
Deflection
Accuracy
(in divisions)
2 mV
10 mV
4.90 to 5.10
5 mV
20 mV
3.92 to 4.08
10 mV
50 mV
4.90 to 5.10
20 mV
100 mV
4.90 to 5.10
50 mV
200 mV
3.92 to 4.08
0.1 V
500 mV
4.90 to 5.10
0.2 V
1 V
4.90 to 5.10
0.5 V
2 V
3.92 to 4.08
1 V
5 V
4.90 to 5.10
2 V
10 V
4.90 to 5.10
5 V
20 V
3.92 to 4.08
10. CH 3 and CH 4 Vertical Deflection Accuracy
a. Set:
Vertical MODE
CH 3 (CH 1 off)
CH 3 & CH 4 VOLTS/DIV
0.1 V
b. Position the trace two graticule lines below the cen
ter horizontal graticule line.
c. Move CH 1 test setup to the CH 3 input.
d. Set calibration generator AMPLITUDE to 0.5 V.
e. C H EC K-the signal amplitude is between 4.90 and
5.10 divisions.
4 -8
REV JUN 1991
Summary of Contents for 2246 1Y
Page 13: ...2246A Service 7062 01 X The 2246 1Y or 2246 Mod A Portable Oscilloscope ...
Page 35: ......
Page 139: ......
Page 185: ......
Page 187: ......
Page 278: ...Figure 9 5b Detailed 2246 1Y or 2246 Mod A block diagram part 2 R E ...
Page 287: ......
Page 304: ...A D D J U N 1991 ...
Page 311: ...2246 1Y and 2246 Mod A Service WAVEFORMS FOR DIAGRAM 1 0 V r l III B f e I l i 6555 39 ...
Page 319: ......
Page 320: ......
Page 321: ...2246 1Y and 2246 Mod A Service WAVEFORMS FOR DIAGRAM 2 SET READOUT CONTROL CCW OFF 6555 40 ...
Page 323: ......
Page 324: ......
Page 326: ...A B C P E F G ...
Page 327: ......
Page 331: ......
Page 337: ......
Page 340: ......
Page 343: ...C l rti ...
Page 347: ...BEV JUNE 1991 7062 18 ...
Page 349: ...2246 1Y and 2246 Mod A Service WAVEFORMS FOR DIAGRAM 6 6555 43 I ...
Page 355: ...I BOARD LOCATION 8K 8J 8H 7H 8H 8H 7J 8D 7H 8H 8J 8J 7K 8K 10D 6L 6K 6L 6K 6G 5H I ...
Page 362: ......
Page 372: ...2246 1Y and 2246 Mod A Service WAVEFORMS FOR DIAGRAM 9 MORE ...
Page 373: ... 6555 72 6557 88 ...
Page 375: ......
Page 380: ...WAVEFORMS FOR DIAGRAM 10 ...
Page 387: ...WAVEFORMS FOR DIAGRAM 11 6555 73 6557 48 ...
Page 394: ...WAVEFORMS FOR DIAGRAM 12 2246 1Y and 2246 Mod A Service i f I S 1 f c i 4 i ...
Page 395: ...WAVEFORMS FOR DIAGRAM 12 cont 2246 1Y and 2246 Mod A Service ...
Page 397: ......
Page 398: ......
Page 400: ...REV JUNE 1991 7062 39 ...
Page 401: ......
Page 403: ......
Page 404: ......
Page 406: ...B 2246 IV AND 2246 MOD A ...
Page 407: ......
Page 408: ......
Page 414: ... _ 5 ...
Page 415: ...2246 1Y 2246 MOD A ...
Page 416: ......
Page 423: ...2246 1Y 2246 MOD A ...
Page 424: ...2246 1Y 2246 MOD A ...
Page 425: ......
Page 433: ......