Performance Check Procedure —2246 1Y and 2246 Mod A Service
g. Move test signal from CH 1 to the CH 2 input.
h. Set leveled sine-wave generator output to pro
duce a 1.0 division display amplitude at 100 MHz.
i.
CHECK—that a stable display can be obtained.
(The Trigger LEVEL control may be adjusted to
improve the display stability.)
j.
Repeat procedure for the CH 3 and CH 4 (turn on
the appropriate Vertical MODE and move the test
signal as required).
5. 100 MHz NOISE REJ Trigger Sensitivity
a. Move test signal to the CH 1 input.
b. Set Vertical MODE to CH 1 (others off).
c. Remove the 2X BNC attenuator from the test sig
nal path.
d. Set Leveled Sine-Wave Generator output for a
2.2 division display amplitude at 100 MHz.
e. CHECK—that the display is stably triggered with
NOISE REJ Trigger CPLG.
6. Single Sweep Mode
a. Set:
Horizontal MODE
A
ASEC/DIV
10 p.s
A/B SELECT
A Trigger
b. Remove the 10X and 2X BNC attenuators from
the test signal path.
c. Set leveled sine-wave generator output to pro
duce a 7.0 division display amplitude at 50 kHz.
d. Add a 10X and a 2X BNC attenuator before the
5 0 - n BNC termination. (Display should stably
trigger with AUTO LEVEL finding the correct trig
ger level setting.)
e. Set:
A Trigger MODE
NORM
CH 1 Input COUPLING
GND
Trigger MODE
SGL SEQ
f.
CHECK—that the Trigger READY LED turns on and
remains on.
g. Set:
A INTEN
3/4 fully CW
CH 1 Input COUPLING
DC (see
CHECK below)
h. CHECK—that the TRIG’ D LED flashes, and the
READY LED turns off after a single sweep and
readout display occurs when the Input COUPLING
switches to DC.
7. Trigger LEVEL Control Range
a. Set:
Trigger MODE
AUTO (not
AUTO LEVEL)
Trigger LEVEL
Fully CCW
A INTEN
For a good
viewing
intensity
b. Remove 10X and 2X BNC attenuators from the
test signal path.
c. Increase leveled sine-wave generator output
level until a stably triggered display is just
obtainable.
d. Set Trigger LEVEL fully CW.
e. Set leveled sine-wave generator output for a sta
ble display (if necessary).
f.
Set CH 1 VOLTS/DIV to 1 V.
g. CHECK—that the CH 1 signal display amplitude is
four divisions or more (p e a k-to -p e a k). Note that
the signal is not triggered.
h. Disconnect the test setup from the 2246 1Y or
2246 Mod A.
4-16
REV OCT 1991
Summary of Contents for 2246 1Y
Page 13: ...2246A Service 7062 01 X The 2246 1Y or 2246 Mod A Portable Oscilloscope ...
Page 35: ......
Page 139: ......
Page 185: ......
Page 187: ......
Page 278: ...Figure 9 5b Detailed 2246 1Y or 2246 Mod A block diagram part 2 R E ...
Page 287: ......
Page 304: ...A D D J U N 1991 ...
Page 311: ...2246 1Y and 2246 Mod A Service WAVEFORMS FOR DIAGRAM 1 0 V r l III B f e I l i 6555 39 ...
Page 319: ......
Page 320: ......
Page 321: ...2246 1Y and 2246 Mod A Service WAVEFORMS FOR DIAGRAM 2 SET READOUT CONTROL CCW OFF 6555 40 ...
Page 323: ......
Page 324: ......
Page 326: ...A B C P E F G ...
Page 327: ......
Page 331: ......
Page 337: ......
Page 340: ......
Page 343: ...C l rti ...
Page 347: ...BEV JUNE 1991 7062 18 ...
Page 349: ...2246 1Y and 2246 Mod A Service WAVEFORMS FOR DIAGRAM 6 6555 43 I ...
Page 355: ...I BOARD LOCATION 8K 8J 8H 7H 8H 8H 7J 8D 7H 8H 8J 8J 7K 8K 10D 6L 6K 6L 6K 6G 5H I ...
Page 362: ......
Page 372: ...2246 1Y and 2246 Mod A Service WAVEFORMS FOR DIAGRAM 9 MORE ...
Page 373: ... 6555 72 6557 88 ...
Page 375: ......
Page 380: ...WAVEFORMS FOR DIAGRAM 10 ...
Page 387: ...WAVEFORMS FOR DIAGRAM 11 6555 73 6557 48 ...
Page 394: ...WAVEFORMS FOR DIAGRAM 12 2246 1Y and 2246 Mod A Service i f I S 1 f c i 4 i ...
Page 395: ...WAVEFORMS FOR DIAGRAM 12 cont 2246 1Y and 2246 Mod A Service ...
Page 397: ......
Page 398: ......
Page 400: ...REV JUNE 1991 7062 39 ...
Page 401: ......
Page 403: ......
Page 404: ......
Page 406: ...B 2246 IV AND 2246 MOD A ...
Page 407: ......
Page 408: ......
Page 414: ... _ 5 ...
Page 415: ...2246 1Y 2246 MOD A ...
Page 416: ......
Page 423: ...2246 1Y 2246 MOD A ...
Page 424: ...2246 1Y 2246 MOD A ...
Page 425: ......
Page 433: ......