Performance Check Procedure—2246 1Y and 2246 Mod A Service
CH1/CH2 VOLTMETER MEASUREMENTS
Equipment Required (See Table 4-1)
Calibration generator
50
O BNC coaxial cable
Leveled sine-wave generator
50
O BNC termination
Function generator
1. DC Volts Accuracy
e.
a. Set:
f.
READOUT (Intensity)
For a viewable
readout
A INTEN
For a viewable
trace
g-
Vertical MODE
CH 1
CH 1 VOLTS/DIV
50 mV
CH 2 INVERT
Off
h.
SCOPE BW
Off
CH 1 Input COUPLING
GND
i.
Horizontal MODE
A
A SEC/DIV
1 ms
j-
A/B SELECT
A Trigger
Trigger MODE
AUTO LEVEL
Trigger CPLG
DC
k.
Trigger SOURCE
VERT
Trigger SLOPE
_r
(positive)
1.
Trigger HOLDOFF
Min
CH1/CH2 VOLTMETER
DC
FOCUS
For best
defined display
m
Horizontal POSITION
12 o ’clock
n.
b. CHECK—ground readout is 0.0 mV ±1.2 mV.
2.
c. Set calibration generator
(PG 506) internal
a.
Square Wave/DC switch to DC.
b.
NOTE
The PG 506 must be removed from the TM
c.
power supply to make the change to dc output
from the generator. Turn the power off before
removing or inserting any plug-in from the TM
d.
power supply
d. Connect the calibration generator Std Ampl out
e.
put to the CH 1 input via a 5 0 - 0 precision BNC
f.
coaxial cable.
Set calibration generator for Std Ampl output of
50 mV dc.
Set:
CH 1 VOLTS/DIV
10 mV
CH 1 Input COUPLING
DC
CHECK—the readout is between 49.0 mV and
51.0 mV.
Set CH 1 VOLTS/DIV to 0.1 V.
Set calibration generator for Std Ampl output
0.5 V.
CHECK—the readout is between 0.495 V and
0.505 V.
Set CH 1 VOLTS/DIV to 1 V.
Set calibration generator for Std Ampl output of 5
V.
CHECK—the readout is between 4.95 V and
5.05 V.
Disconnect Std Ampl signal from the CH 1 input.
DC Volts Normal Mode Rejection Ratio
Set SEC/DIV to 5 ms.
Connect function generator (FG 502) output to
the CH 1 input via a 5 0 - 0 BNC coaxial cable and
BNC coupling capacitor.
Set function generator for a six-division sine-
wave display amplitude at 50 Hz (with CH 1
VOLTS/DIV at 1 V ).
Set CH 1 VOLTS/DIV to 0.2 V.
CHECK—that the readout is less than ±0.018 V.
Disconnect the function generator signal from the
2246 1Y or 2246 Mod A.
REV OCT 1991
4-25
Summary of Contents for 2246 1Y
Page 13: ...2246A Service 7062 01 X The 2246 1Y or 2246 Mod A Portable Oscilloscope ...
Page 35: ......
Page 139: ......
Page 185: ......
Page 187: ......
Page 278: ...Figure 9 5b Detailed 2246 1Y or 2246 Mod A block diagram part 2 R E ...
Page 287: ......
Page 304: ...A D D J U N 1991 ...
Page 311: ...2246 1Y and 2246 Mod A Service WAVEFORMS FOR DIAGRAM 1 0 V r l III B f e I l i 6555 39 ...
Page 319: ......
Page 320: ......
Page 321: ...2246 1Y and 2246 Mod A Service WAVEFORMS FOR DIAGRAM 2 SET READOUT CONTROL CCW OFF 6555 40 ...
Page 323: ......
Page 324: ......
Page 326: ...A B C P E F G ...
Page 327: ......
Page 331: ......
Page 337: ......
Page 340: ......
Page 343: ...C l rti ...
Page 347: ...BEV JUNE 1991 7062 18 ...
Page 349: ...2246 1Y and 2246 Mod A Service WAVEFORMS FOR DIAGRAM 6 6555 43 I ...
Page 355: ...I BOARD LOCATION 8K 8J 8H 7H 8H 8H 7J 8D 7H 8H 8J 8J 7K 8K 10D 6L 6K 6L 6K 6G 5H I ...
Page 362: ......
Page 372: ...2246 1Y and 2246 Mod A Service WAVEFORMS FOR DIAGRAM 9 MORE ...
Page 373: ... 6555 72 6557 88 ...
Page 375: ......
Page 380: ...WAVEFORMS FOR DIAGRAM 10 ...
Page 387: ...WAVEFORMS FOR DIAGRAM 11 6555 73 6557 48 ...
Page 394: ...WAVEFORMS FOR DIAGRAM 12 2246 1Y and 2246 Mod A Service i f I S 1 f c i 4 i ...
Page 395: ...WAVEFORMS FOR DIAGRAM 12 cont 2246 1Y and 2246 Mod A Service ...
Page 397: ......
Page 398: ......
Page 400: ...REV JUNE 1991 7062 39 ...
Page 401: ......
Page 403: ......
Page 404: ......
Page 406: ...B 2246 IV AND 2246 MOD A ...
Page 407: ......
Page 408: ......
Page 414: ... _ 5 ...
Page 415: ...2246 1Y 2246 MOD A ...
Page 416: ......
Page 423: ...2246 1Y 2246 MOD A ...
Page 424: ...2246 1Y 2246 MOD A ...
Page 425: ......
Page 433: ......