Chapter 9 – Temporary Basal Rate
9.1
Setting a Temp Rate
A Temp Rate is used to increase or
decrease (by percentage) the current
basal rate for a period of time. This fea-
ture can be helpful for situations such
as exercise or illness.
When you enter the Temp Rate screen,
the default values are 100% (current
basal rate) and a Duration of 0:15
min. The Temp Rate can be set from a
minimum of 0% of current basal rate to
a maximum of 250% of current basal
rate in increments of 1%.
Duration can be set from a minimum of
15 minutes to a maximum of 72 hours
in increments of 1 minute.
If you program a Temp Rate greater
than 0% but less than the minimum
allowable basal rate of 0.1 units/hr, you
will be notified that the selected rate
is too low and that it will be set to the
minimum allowable rate for delivery.
If you program a Temp Rate more than
the maximum allowable basal rate of
15 units/hr, you will be notified that the
selected rate is too high and that it will
NOTE
If a Temp Rate is active, the status of
insulin delivery displays an orange “T”. If
a Temp Rate of 0% is active, the status of
insulin delivery displays a red “T”.
8.
The Screen Lock screen will be
displayed with the icon indicating
a Temp Rate is active.
be set to the maximum allowable rate
for delivery.
1.
From the Home Screen,
tap
OPTIONS
.
2.
Tap
Temp Rate
.
3.
Tap
Temp Rate
.
4.
Using the onscreen keypad enter
desired percentage. Tap
DONE
.
NOTE
Current rate is 100%. An increase is
greater than 100% and decrease is less
than 100%.
5.
Tap
Duration
. Using the onscreen
keypad enter desired length of
time for Temp Rate. Tap
DONE
.
NOTE
To see the actual units to be delivered tap
View Units.
6.
Verify settings and tap
START
.
7.
The “TEMP RATE STARTED”
splash screen is displayed.
Summary of Contents for T:Slim
Page 2: ......
Page 16: ...Section 1 Welcome...
Page 17: ...Chapter 1 About this User Guide...
Page 27: ...Chapter 2 Introduction...
Page 35: ...Chapter 3 t slim Pump Overview...
Page 54: ...Section 2 Key Pump Features...
Page 55: ...Chapter 4 Getting Started...
Page 61: ...Chapter 5 Infusion Site Care and Loading Cartridge...
Page 75: ...Chapter 6 Personal Profile...
Page 85: ...Chapter 7 Bolus...
Page 93: ...Chapter 8 Stop Resume Insulin...
Page 96: ...Section 3 Additional Pump Operations...
Page 97: ...Chapter 9 Temporary Basal Rate...
Page 101: ...Chapter 10 Quick Bolus...
Page 105: ...Chapter 11 t slim Pump Settings...
Page 109: ...Chapter 12 t slim Pump Info and History...
Page 111: ...Chapter 13 t slim Pump Reminders...
Page 117: ...Chapter 14 User Settable Alerts and Alarms...
Page 120: ...Section 4 t slim Pump Safety Alerts and Alarms...
Page 121: ...Chapter 15 t slim Pump Alerts...
Page 143: ...Chapter 16 t slim Pump Alarms...
Page 157: ...Chapter 17 t slim Pump Malfunction...
Page 160: ...Section 5 System Care and Maintenance...
Page 161: ...Chapter 18 Taking Care of Your t slim System...
Page 164: ...Section 6 System Technical Specifications...
Page 165: ...Chapter 19 Technical Specifications...
Page 187: ......