Key Pump Featur
es
2
Chapter 6 – Personal Profile
•
Correction Factor (range: 1 unit:1
mg/dL – 1 unit:600 mg/dL)
•
Carb Ratio (range: 1 unit:1 gram –
1 unit:300 grams)
•
Target BG (range: 70 mg/dL –
250 mg/dL)
NOTE
Below a carb ratio of 1:10, increments can
be entered in 0.1 g. For example a carb
ratio of 1:8.2 can be programmed.
In addition, you can set any or all of
the following Bolus Settings:
•
Carbs (on indicates entering grams
of Carb; off indicates entering units
of insulin)
•
Insulin Duration (how long a bolus
lowers your BG)
•
Max Bolus (the maximum amount
for a single bolus)
The default settings and ranges for
Bolus Settings are as follows:
•
Carbs (default: off if no Carb
Ratio is defined)
•
Insulin Duration (default: 5 hrs;
range: 2–8 hrs)
•
Max Bolus (default: 10 units;
range: 1–25 units)
Insulin Duration and Insulin on
Board (IOB)
Your t:slim Pump remembers how
much insulin you have taken from
previous boluses. It does this by relying
on the Insulin Duration. The Insulin Du-
ration reflects the amount of time that
insulin is actively lowering your BG.
Insulin Duration in your t:slim Pump is
set from the factory at 5 hours but can
be changed from 2 to 8 hours. Please
consult your healthcare provider to
accurately set your Insulin Duration.
While the Insulin Duration setting
reflects how long insulin from previ-
ous boluses lowers your BG, the IOB
feature reflects how much insulin is
remaining in your body from previous
boluses. IOB is always displayed on
the Home Screen and is used in bolus
delivery calculations when applicable.
When a BG is entered during bolus
programming, your t:slim Pump will
consider any active IOB and calculate
an adjusted bolus if necessary.
NOTE
Although you do not need to define every
setting, some pump features require certain
settings to be defined and activated. During
profile setup, your pump prompts you to set up
any required settings before you can continue.
1.
From the Home Screen,
tap
OPTIONS
.
2.
Tap
My Pump
.
3.
Tap
Personal Profiles
.
4.
Tap
NEW
to create a new profile.
5.
Using the onscreen keypad,
enter a profile name (up to 16
characters) and tap
SAVE
.
Summary of Contents for T:Slim
Page 2: ......
Page 16: ...Section 1 Welcome...
Page 17: ...Chapter 1 About this User Guide...
Page 27: ...Chapter 2 Introduction...
Page 35: ...Chapter 3 t slim Pump Overview...
Page 54: ...Section 2 Key Pump Features...
Page 55: ...Chapter 4 Getting Started...
Page 61: ...Chapter 5 Infusion Site Care and Loading Cartridge...
Page 75: ...Chapter 6 Personal Profile...
Page 85: ...Chapter 7 Bolus...
Page 93: ...Chapter 8 Stop Resume Insulin...
Page 96: ...Section 3 Additional Pump Operations...
Page 97: ...Chapter 9 Temporary Basal Rate...
Page 101: ...Chapter 10 Quick Bolus...
Page 105: ...Chapter 11 t slim Pump Settings...
Page 109: ...Chapter 12 t slim Pump Info and History...
Page 111: ...Chapter 13 t slim Pump Reminders...
Page 117: ...Chapter 14 User Settable Alerts and Alarms...
Page 120: ...Section 4 t slim Pump Safety Alerts and Alarms...
Page 121: ...Chapter 15 t slim Pump Alerts...
Page 143: ...Chapter 16 t slim Pump Alarms...
Page 157: ...Chapter 17 t slim Pump Malfunction...
Page 160: ...Section 5 System Care and Maintenance...
Page 161: ...Chapter 18 Taking Care of Your t slim System...
Page 164: ...Section 6 System Technical Specifications...
Page 165: ...Chapter 19 Technical Specifications...
Page 187: ......