WARNING
When using the Car Power USB Adapter, the
charger must be connected to an isolated,
battery powered 12 Volt system, such as an
automobile. Connecting the DC vehicle adapter
charger to 12 Volt DC that is generated by a
power supply connected to alternating current
(AC) mains is prohibited.
To charge the pump using a USB port
on a Personal Computer (PC):
Ensure that the PC complies with
the IEC 60950-1 (or equivalent)
safety standard.
1.
Plug the included USB cable into
your computer.
2.
Plug the other end of the cable
into the micro USB port on the
pump.
NOTE
Before using a Mac or PC to charge the t:slim
Pump, it is recommended that a driver be
installed on the computer by downloading the
t:connect Uploader Software from our website
at www.tandemdiabetes.com. This will also
allow communication between the pump, the
PC, and the t:connect Application. Depending
on your computer, charging time will vary.
The pump will display a CONNECTION ERROR
ALERT message if it is not charging properly.
When you charge the t:slim Pump,
you will notice the following:
•
The screen illuminates
•
An audible alert
•
The LED (edge around the
Screen On/Quick Bolus Button)
blinks green
•
A vibrating alert
•
A charge symbol (lightning bolt) on
the battery level indicator appears
PRECAUTION
If any of these conditions fail to occur, do
NOT use the t:slim Pump, as this may be an
indication that the pump’s alarms, alerts, and
other built-in safety features are not working
properly. If the t:slim Pump is connected to
your body, disconnect the set from your body
and contact Tandem Diabetes Care Customer
Technical Support at 1-877-801-6901.
Charging Tips
Tandem Diabetes Care, Inc. rec-
ommends periodically checking the
battery level indicator, charging the
pump for a short period of time every
day (10–15 minutes), and also avoiding
frequent full discharges.
NOTE
If the battery is fully discharged, the screen will
not power on immediately when connected to
a charging source. The LED around the Screen
On/Quick Bolus Button will blink green every 2
seconds until there is enough charge to power
on the touch screen.
Summary of Contents for T:Slim
Page 2: ......
Page 16: ...Section 1 Welcome...
Page 17: ...Chapter 1 About this User Guide...
Page 27: ...Chapter 2 Introduction...
Page 35: ...Chapter 3 t slim Pump Overview...
Page 54: ...Section 2 Key Pump Features...
Page 55: ...Chapter 4 Getting Started...
Page 61: ...Chapter 5 Infusion Site Care and Loading Cartridge...
Page 75: ...Chapter 6 Personal Profile...
Page 85: ...Chapter 7 Bolus...
Page 93: ...Chapter 8 Stop Resume Insulin...
Page 96: ...Section 3 Additional Pump Operations...
Page 97: ...Chapter 9 Temporary Basal Rate...
Page 101: ...Chapter 10 Quick Bolus...
Page 105: ...Chapter 11 t slim Pump Settings...
Page 109: ...Chapter 12 t slim Pump Info and History...
Page 111: ...Chapter 13 t slim Pump Reminders...
Page 117: ...Chapter 14 User Settable Alerts and Alarms...
Page 120: ...Section 4 t slim Pump Safety Alerts and Alarms...
Page 121: ...Chapter 15 t slim Pump Alerts...
Page 143: ...Chapter 16 t slim Pump Alarms...
Page 157: ...Chapter 17 t slim Pump Malfunction...
Page 160: ...Section 5 System Care and Maintenance...
Page 161: ...Chapter 18 Taking Care of Your t slim System...
Page 164: ...Section 6 System Technical Specifications...
Page 165: ...Chapter 19 Technical Specifications...
Page 187: ......