Customer Technical Support:
1-877-801-6901
113
Additional Pump Operations
3
Chapter 13 – t:slim Pump Reminders
6.
Low BG is set to on; to turn off,
tap
LOW BG
.
a.
Tap
Remind Me Below
and us-
ing the onscreen keypad, enter a
Low BG value (from 70–120 mg/
dL) that you want to trigger the
reminder, then tap
DONE
.
b.
Tap
Remind Me After
and using
the onscreen keypad, enter the
time (from 10–20 min), then tap
DONE
.
c.
Tap
SAVE
when all changes
are complete.
d.
Tap
Tandem Logo
to return to
the Home Screen.
To Respond to the Low
BG Reminder
To clear the reminder, tap
CLOSE
and then check BG using your blood
glucose meter.
13.2 High BG Reminder
The High BG Reminder prompts you to
re-test your BG after a high BG value is
entered. When you turn this reminder
on, you need to set a high BG value
that triggers the reminder, as well as
how much time should pass before the
reminder occurs.
The default for this reminder is pre-set
to off. If on, Remind Me Above 200
mg/dL, and Remind Me After 120 min,
but you can set these values from
150–300 mg/dL and 1–3 hrs.
1.
From the Home Screen,
tap
OPTIONS
.
2.
Tap
My Pump
.
3.
Tap
Alert Settings
.
4.
Tap
Pump Reminders
.
5.
Tap
High BG
.
NOTE
If not first-time setup, previous values
appear instead of “Press to Turn On”.
6.
High BG is set to on; to turn off,
tap
High BG
.
a.
Tap
Remind Me Above
and
using the onscreen keypad, enter
a High BG value (from 150–300
mg/dL) that you want to trigger
the reminder, then tap
DONE
.
b.
Tap
Remind Me After
and using
the onscreen keypad, enter the
time (from 1–3 hours), then tap
DONE
.
c.
Tap
SAVE
when all changes
are complete.
7.
Tap
Tandem Logo
to return to the
Home Screen.
To Respond to the High
BG Reminder
To clear the reminder tap
CLOSE
and then check BG using your blood
glucose meter.
Summary of Contents for T:Slim
Page 2: ......
Page 16: ...Section 1 Welcome...
Page 17: ...Chapter 1 About this User Guide...
Page 27: ...Chapter 2 Introduction...
Page 35: ...Chapter 3 t slim Pump Overview...
Page 54: ...Section 2 Key Pump Features...
Page 55: ...Chapter 4 Getting Started...
Page 61: ...Chapter 5 Infusion Site Care and Loading Cartridge...
Page 75: ...Chapter 6 Personal Profile...
Page 85: ...Chapter 7 Bolus...
Page 93: ...Chapter 8 Stop Resume Insulin...
Page 96: ...Section 3 Additional Pump Operations...
Page 97: ...Chapter 9 Temporary Basal Rate...
Page 101: ...Chapter 10 Quick Bolus...
Page 105: ...Chapter 11 t slim Pump Settings...
Page 109: ...Chapter 12 t slim Pump Info and History...
Page 111: ...Chapter 13 t slim Pump Reminders...
Page 117: ...Chapter 14 User Settable Alerts and Alarms...
Page 120: ...Section 4 t slim Pump Safety Alerts and Alarms...
Page 121: ...Chapter 15 t slim Pump Alerts...
Page 143: ...Chapter 16 t slim Pump Alarms...
Page 157: ...Chapter 17 t slim Pump Malfunction...
Page 160: ...Section 5 System Care and Maintenance...
Page 161: ...Chapter 18 Taking Care of Your t slim System...
Page 164: ...Section 6 System Technical Specifications...
Page 165: ...Chapter 19 Technical Specifications...
Page 187: ......