Customer Technical Support:
1-877-801-6901
179
System T
echnical Specifications
6
Chapter 19 – Technical Specifications
Recommended Separation Distances Between Portable and Mobile
RF Communications Equipment and the t:slim System
The t:slim System is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of
the t:slim System can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the System as recommended below, according to the maximum output power of the communications equipment.
Rated Maximum
Output Power of
Transmitter in Watts
Separation Distance According to Frequency of Transmitter
inches or feet
(m)
150 kHz to 80 MHz
(d
in meters
=
0.35
√
P)
80 MHz to 800 MHz
(d
in meters
=
0.12
√
P)
800 MHz to 2.5 GHz
(d
in meters
=
0.23
√
P)
0.01
1.6 in (0.04)
0.12 in (0.012)
0.9 in (0.023)
0.1
4.3 in (0.11)
1.5 in (0.038)
2.9 in (0.073)
1
1.1 ft (0.35)
4.7 in (0.120)
9.0 in (0.23)
10
3.6 ft (1.11)
14.9 in (0.379)
2.39 ft (0.727)
100
11.5 ft (3.50)
3.94 ft (1.20)
7.54 ft (2.3)
For transmitters rated at a maximum output power not listed above, the recommended separation distance (d) in meters (m) can be determined using
the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects,
and people.
Summary of Contents for T:Slim
Page 2: ......
Page 16: ...Section 1 Welcome...
Page 17: ...Chapter 1 About this User Guide...
Page 27: ...Chapter 2 Introduction...
Page 35: ...Chapter 3 t slim Pump Overview...
Page 54: ...Section 2 Key Pump Features...
Page 55: ...Chapter 4 Getting Started...
Page 61: ...Chapter 5 Infusion Site Care and Loading Cartridge...
Page 75: ...Chapter 6 Personal Profile...
Page 85: ...Chapter 7 Bolus...
Page 93: ...Chapter 8 Stop Resume Insulin...
Page 96: ...Section 3 Additional Pump Operations...
Page 97: ...Chapter 9 Temporary Basal Rate...
Page 101: ...Chapter 10 Quick Bolus...
Page 105: ...Chapter 11 t slim Pump Settings...
Page 109: ...Chapter 12 t slim Pump Info and History...
Page 111: ...Chapter 13 t slim Pump Reminders...
Page 117: ...Chapter 14 User Settable Alerts and Alarms...
Page 120: ...Section 4 t slim Pump Safety Alerts and Alarms...
Page 121: ...Chapter 15 t slim Pump Alerts...
Page 143: ...Chapter 16 t slim Pump Alarms...
Page 157: ...Chapter 17 t slim Pump Malfunction...
Page 160: ...Section 5 System Care and Maintenance...
Page 161: ...Chapter 18 Taking Care of Your t slim System...
Page 164: ...Section 6 System Technical Specifications...
Page 165: ...Chapter 19 Technical Specifications...
Page 187: ......