Chapter 5 – Infusion Site Care and Loading Cartridge
5.2
Cartridge Instructions
for Use
For complete cartridge labeling, consult
the Cartridge Instructions for Use
included in the t:slim Cartridge box.
5.3
Filling and Loading a
t:slim Cartridge
This section describes how to fill the
cartridge with insulin and load the
cartridge into your t:slim Pump. The
single-use disposable cartridge can
hold up to 300 units (3.0 mL) of insulin.
PRECAUTION
Replace the cartridge every 48 hours if using
Humalog
®
; every 72 hours if using Novolog
®
.
PRECAUTION
Inspect the cartridge to ensure that there is no
damage. Never use a cartridge that is damaged
(cracks, dislodged tip connectors, etc.). If
there is damage, call Tandem Diabetes Care
Customer Technical Support at 1-877-801-
6901 to request a replacement cartridge.
PRECAUTION
Before you begin, verify that the infusion set is
NOT connected to your body.
Before you begin, make sure you
have the following items:
•
1 unopened cartridge
•
3.0 mL syringe and fill needle
•
Vial of Humalog
®
or NovoLog
®
insulin
•
Alcohol prep swab
•
1 new infusion set
•
Infusion set Instructions for Use
NOTE
Do NOT remove the used cartridge from the
pump during the load process until prompted
on the t:slim Pump screen.
The illustration identifies the
luer-lock and insulin fill port used in
the cartridge filling process.
Luer-Lock
Tail of the
Cartridge
Insulin
Fill Port
Summary of Contents for T:Slim
Page 2: ......
Page 16: ...Section 1 Welcome...
Page 17: ...Chapter 1 About this User Guide...
Page 27: ...Chapter 2 Introduction...
Page 35: ...Chapter 3 t slim Pump Overview...
Page 54: ...Section 2 Key Pump Features...
Page 55: ...Chapter 4 Getting Started...
Page 61: ...Chapter 5 Infusion Site Care and Loading Cartridge...
Page 75: ...Chapter 6 Personal Profile...
Page 85: ...Chapter 7 Bolus...
Page 93: ...Chapter 8 Stop Resume Insulin...
Page 96: ...Section 3 Additional Pump Operations...
Page 97: ...Chapter 9 Temporary Basal Rate...
Page 101: ...Chapter 10 Quick Bolus...
Page 105: ...Chapter 11 t slim Pump Settings...
Page 109: ...Chapter 12 t slim Pump Info and History...
Page 111: ...Chapter 13 t slim Pump Reminders...
Page 117: ...Chapter 14 User Settable Alerts and Alarms...
Page 120: ...Section 4 t slim Pump Safety Alerts and Alarms...
Page 121: ...Chapter 15 t slim Pump Alerts...
Page 143: ...Chapter 16 t slim Pump Alarms...
Page 157: ...Chapter 17 t slim Pump Malfunction...
Page 160: ...Section 5 System Care and Maintenance...
Page 161: ...Chapter 18 Taking Care of Your t slim System...
Page 164: ...Section 6 System Technical Specifications...
Page 165: ...Chapter 19 Technical Specifications...
Page 187: ......