173
System T
echnical Specifications
6
Chapter 19 – Technical Specifications
…table continued from previous page
Option/Setting Type
Option/Setting Details
Quick Bolus Increments
When set to units: 0.5, 1, 2, 5 units (default is 0.5 units);
or when set to grams/carbs: 2, 5, 10, 15 grams (default is 2 g)
Maximum Extended Bolus Time
8 hours
Maximum Bolus Size
25 units
Low Reservoir Volume Indicator
Status indicator visible on Home Screen; Low Insulin Alert is user adjustable from 10–40 units (default is 20 units).
Auto-Off Alarm
On or Off (default is On); user-adjustable (5–24 hours; default is 12 hours, which you can change when option is set to On).
History Storage
Up to 90 days of data (11,000 events).
Language
English
Feature Lock
Blocks access to insulin delivery screens and pump setting screens (default is Off).
Screen Lock
Protects from unintentional taps.
Site Reminder
Prompts user to change infusion set. Can be set for 1–3 days at a time selected by user (default is Off).
Missed Meal Bolus Reminder
Prompts user if a bolus has not occurred during the period of time the reminder is set for.
4 reminders available (default is Off).
After Bolus BG Reminder
Prompts user to test BG at a selected time period after a Bolus has been delivered.
Can be set between 1-3 hours (default is Off).
High BG Reminder
Prompts user to retest BG after a High BG has been entered. User selects High BG value and time for reminder.
(default is Off).
Low BG Reminder
Prompts user to retest BG after a Low BG has been entered. User selects Low BG value and time for reminder.
(default is Off).
Summary of Contents for T:Slim
Page 2: ......
Page 16: ...Section 1 Welcome...
Page 17: ...Chapter 1 About this User Guide...
Page 27: ...Chapter 2 Introduction...
Page 35: ...Chapter 3 t slim Pump Overview...
Page 54: ...Section 2 Key Pump Features...
Page 55: ...Chapter 4 Getting Started...
Page 61: ...Chapter 5 Infusion Site Care and Loading Cartridge...
Page 75: ...Chapter 6 Personal Profile...
Page 85: ...Chapter 7 Bolus...
Page 93: ...Chapter 8 Stop Resume Insulin...
Page 96: ...Section 3 Additional Pump Operations...
Page 97: ...Chapter 9 Temporary Basal Rate...
Page 101: ...Chapter 10 Quick Bolus...
Page 105: ...Chapter 11 t slim Pump Settings...
Page 109: ...Chapter 12 t slim Pump Info and History...
Page 111: ...Chapter 13 t slim Pump Reminders...
Page 117: ...Chapter 14 User Settable Alerts and Alarms...
Page 120: ...Section 4 t slim Pump Safety Alerts and Alarms...
Page 121: ...Chapter 15 t slim Pump Alerts...
Page 143: ...Chapter 16 t slim Pump Alarms...
Page 157: ...Chapter 17 t slim Pump Malfunction...
Page 160: ...Section 5 System Care and Maintenance...
Page 161: ...Chapter 18 Taking Care of Your t slim System...
Page 164: ...Section 6 System Technical Specifications...
Page 165: ...Chapter 19 Technical Specifications...
Page 187: ......