Customer Technical Support:
1-877-801-6901
168
Chapter 19 – Technical Specifications
…table continued from previous page
Specification Type
Specification Details
Statement of Meaning of the Claimed
Ingress Protection (IP) Classification
IPX7: watertight to a depth of 3 feet for up to 30 minutes
Internal System Battery Life
4 years minimum under normal use conditions
Typical Operating Time when System
is Operating at Intermediate Rate
During normal use, the intermediate rate is 2 units/hr; battery charge can be reasonably expected to last up to
7 days from a fully charged state to a totally discharged state
Handling of Over-Infusion or
Under-Infusion
The method of delivery isolates the insulin chamber from the patient and the software performs frequent
monitoring of system status. Multiple software monitors provide redundant protection against unsafe
conditions.
Over-infusion is mitigated by continuous self-tests, layering of redundancies and confirmations, and numerous
other safeguard alarms. Users are required to review and confirm the details of all bolus deliveries, basal rates,
and temp rates to ensure certainty before initiating a delivery. In addition, once bolus deliveries are confirmed,
the user is given 5 seconds to cancel the delivery before it is started. An optional Auto-Off alarm triggers when
the user has not interacted with the pump’s user interface for a pre-defined period of time.
Under-infusion is mitigated by occlusion detection and blood glucose monitoring as blood glucose entries are
recorded. Users are prompted to treat high blood glucose conditions with a correction bolus.
Bolus Volume at Release of Occlusion
0.4–0.6 units at 2 units per hour with Unomedical Comfort
™
(110cm) infusion set.
Alarm Signal Sound Pressure Range
Minimum sound pressure level of 45 dBA at 1 meter
Summary of Contents for T:Slim
Page 2: ......
Page 16: ...Section 1 Welcome...
Page 17: ...Chapter 1 About this User Guide...
Page 27: ...Chapter 2 Introduction...
Page 35: ...Chapter 3 t slim Pump Overview...
Page 54: ...Section 2 Key Pump Features...
Page 55: ...Chapter 4 Getting Started...
Page 61: ...Chapter 5 Infusion Site Care and Loading Cartridge...
Page 75: ...Chapter 6 Personal Profile...
Page 85: ...Chapter 7 Bolus...
Page 93: ...Chapter 8 Stop Resume Insulin...
Page 96: ...Section 3 Additional Pump Operations...
Page 97: ...Chapter 9 Temporary Basal Rate...
Page 101: ...Chapter 10 Quick Bolus...
Page 105: ...Chapter 11 t slim Pump Settings...
Page 109: ...Chapter 12 t slim Pump Info and History...
Page 111: ...Chapter 13 t slim Pump Reminders...
Page 117: ...Chapter 14 User Settable Alerts and Alarms...
Page 120: ...Section 4 t slim Pump Safety Alerts and Alarms...
Page 121: ...Chapter 15 t slim Pump Alerts...
Page 143: ...Chapter 16 t slim Pump Alarms...
Page 157: ...Chapter 17 t slim Pump Malfunction...
Page 160: ...Section 5 System Care and Maintenance...
Page 161: ...Chapter 18 Taking Care of Your t slim System...
Page 164: ...Section 6 System Technical Specifications...
Page 165: ...Chapter 19 Technical Specifications...
Page 187: ......