Chapter 7 – Bolus
7.1
Bolus Overview
A bolus is a quick dose of insulin that is
usually delivered to cover food eaten or
to correct a high BG.
Your t:slim Pump offers you the ability
to deliver different boluses to cover
carbohydrate intake (food bolus) and
bring your BG back to target (correction
bolus). Food and correction boluses
can also be programed together.
If Carbohydrates is turned on in your
active personal profile, you will enter
grams of carbohydrate and the bolus
will be calculated using your Carb
Ratio.
You can always override the calculated
bolus by tapping the area above “units”
located at the top of the bolus screen
between BACK and NEXT. The override
function is always active.
If Carbohydrates is turned off in your
active personal profile, you will enter
units of insulin to request the bolus.
NOTE
When attempting to deliver a bolus that exceeds
the amount of insulin remaining in the cartridge,
a message screen appears indicating there
is not enough insulin to deliver the requested
bolus. You are required to enter a new amount.
WARNING
Do not deliver a bolus until you have reviewed
the calculated bolus amount on the pump
display. If you dose an insulin amount that is too
high or too low, this could lead to a very low or
very high BG level. You can always adjust the
insulin units up or down before you decide to
deliver your bolus.
PRECAUTION
»
Before delivering insulin, always check your
pump and infusion set for damage, leaks, or
kinks in the tubing, as these problems could
restrict or stop insulin delivery and possibly
result in under-infusion.
»
Before delivering insulin, check your t:slim
Pump’s personal settings to ensure they are
correct.
NOTE
If the max bolus is set at 25 units and the
requested bolus is greater than 25 units, after
the bolus is delivered a reminder screen will
appear. The option of delivering the remaining
amount of the bolus of up to an additional 25
units will be given. Refer to Max Bolus Alert in
Chapter 15.9.
Summary of Contents for T:Slim
Page 2: ......
Page 16: ...Section 1 Welcome...
Page 17: ...Chapter 1 About this User Guide...
Page 27: ...Chapter 2 Introduction...
Page 35: ...Chapter 3 t slim Pump Overview...
Page 54: ...Section 2 Key Pump Features...
Page 55: ...Chapter 4 Getting Started...
Page 61: ...Chapter 5 Infusion Site Care and Loading Cartridge...
Page 75: ...Chapter 6 Personal Profile...
Page 85: ...Chapter 7 Bolus...
Page 93: ...Chapter 8 Stop Resume Insulin...
Page 96: ...Section 3 Additional Pump Operations...
Page 97: ...Chapter 9 Temporary Basal Rate...
Page 101: ...Chapter 10 Quick Bolus...
Page 105: ...Chapter 11 t slim Pump Settings...
Page 109: ...Chapter 12 t slim Pump Info and History...
Page 111: ...Chapter 13 t slim Pump Reminders...
Page 117: ...Chapter 14 User Settable Alerts and Alarms...
Page 120: ...Section 4 t slim Pump Safety Alerts and Alarms...
Page 121: ...Chapter 15 t slim Pump Alerts...
Page 143: ...Chapter 16 t slim Pump Alarms...
Page 157: ...Chapter 17 t slim Pump Malfunction...
Page 160: ...Section 5 System Care and Maintenance...
Page 161: ...Chapter 18 Taking Care of Your t slim System...
Page 164: ...Section 6 System Technical Specifications...
Page 165: ...Chapter 19 Technical Specifications...
Page 187: ......