Chapter 19 – Technical Specifications
19.6 Electromagnetic Immunity
The t:slim System is intended for use in the electromagnetic environment specified below. Always make sure that the System is
used in such an environment.
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment –
Guidance
Electrostatic Discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidity
should be at least 30%.
Electrical Fast Transient/burst
IEC 61000-4-4
± 2 kV for power supply lines
± 1 kV for input/ output lines
± 2 kV for power supply lines
± 1 kV for input/output lines
Mains power quality should be that of a typical commercial
or hospital environment.
Surge
IEC 61000-4-5
± 1 kV differential mode
± 2 kV common mode
± 1 kV differential mode
± 2 kV common mode
Mains power quality should be that of a typical commercial
or hospital environment.
Voltage Dips, Short Interruptions,
and Voltage Variations on Power
Supply Input Lines IEC 61000-4-11
<5% Ur (>95% dip in Ur) for 0,5 cycle
40% Ur (60% dip in Ur) for 5 cycles
70% Ur (30% dip in Ur) for 25 cycles
<5% Ur (>95% dip in Ur) for 5 sec
<5% Ur (>95% dip in Ur) for 0,5 cycle
40% Ur (60% dip in Ur) for 5 cycles
70% Ur (30% dip in Ur) for 25 cycles
<5% Ur (>95% dip in Ur) for 5 sec
Mains power quality should be that of a typical commercial
or hospital environment. If the user of the pump requires
continued operation during power mains interruptions,
it is recommended that the pump be powered from an
uninterruptible power supply or a battery.
NOTE: Ur is the a.c. mains voltage prior to application of
the test level.
Power Frequency (50/60 Hz)
Magnetic Field IEC 61000-4-8
3 A/m
400 A/m (IEC 60601-2-24)t
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial
or hospital environment.
table continued on next page …
Summary of Contents for T:Slim
Page 2: ......
Page 16: ...Section 1 Welcome...
Page 17: ...Chapter 1 About this User Guide...
Page 27: ...Chapter 2 Introduction...
Page 35: ...Chapter 3 t slim Pump Overview...
Page 54: ...Section 2 Key Pump Features...
Page 55: ...Chapter 4 Getting Started...
Page 61: ...Chapter 5 Infusion Site Care and Loading Cartridge...
Page 75: ...Chapter 6 Personal Profile...
Page 85: ...Chapter 7 Bolus...
Page 93: ...Chapter 8 Stop Resume Insulin...
Page 96: ...Section 3 Additional Pump Operations...
Page 97: ...Chapter 9 Temporary Basal Rate...
Page 101: ...Chapter 10 Quick Bolus...
Page 105: ...Chapter 11 t slim Pump Settings...
Page 109: ...Chapter 12 t slim Pump Info and History...
Page 111: ...Chapter 13 t slim Pump Reminders...
Page 117: ...Chapter 14 User Settable Alerts and Alarms...
Page 120: ...Section 4 t slim Pump Safety Alerts and Alarms...
Page 121: ...Chapter 15 t slim Pump Alerts...
Page 143: ...Chapter 16 t slim Pump Alarms...
Page 157: ...Chapter 17 t slim Pump Malfunction...
Page 160: ...Section 5 System Care and Maintenance...
Page 161: ...Chapter 18 Taking Care of Your t slim System...
Page 164: ...Section 6 System Technical Specifications...
Page 165: ...Chapter 19 Technical Specifications...
Page 187: ......