Chapter 5 – Infusion Site Care and Loading Cartridge
NOTE
The following are approximate amounts of
insulin to fill different tubing lengths:
»
15–20 units for 23 inch tubing
»
20–25 units for 32 inch tubing
»
25–30 units for 42 inch tubing
4.
Tap
STOP
after 3 drops of insulin
are seen at the end of the infusion
set tubing.
Z
Z
“Stopping Fill” screen is displayed.
Z
Z
“Detecting Insulin” screen
is displayed.
5.
Verify that drops are seen and
tap
DONE
.
•
If you do not see drops, tap
FILL
.
The Fill Tubing screen appears,
repeat steps 3–5 until you see 3
drops of insulin at the end of the
tubing.
•
If you did not tap STOP, the “Max
fill amount reached!” screen will
appear. Do one of the following:
a.
If you are finished filling the tub-
ing, tap
DONE
.
b.
If you want to fill the tubing with
more than 30 units, tap
FILL
to
go back to the Fill Tubing screen.
6.
Fill Tubing is complete.
NOTE
After tubing fill is complete, when the pump
returns to the Home Screen, an estimate
of how much insulin is in the cartridge is
displayed in the upper right portion of the
screen. You will see one of the following on
the screen:
»
+ 40 u More than 40 units
detected in the cartridge
»
+ 60 u More than 60 units
detected in the cartridge
»
+ 120 u More than 120 units
detected in the cartridge
»
+ 180 u More than 180 units
detected in the cartridge
»
+ 240 u More than 240 units
detected in the cartridge
After 10 units are delivered, an actual
number of units remaining in the cartridge
will be displayed on the Home Screen
The amount of insulin remaining displayed
on the Home Screen will decrease 5 units
at a time (for example, you will see 140,
135, 130, 125). When less than 40 units
remain, it will begin decreasing 1 unit at a
time (for example, you will see 40, 39, 38,
37) until there is 1 unit remaining.
Summary of Contents for T:Slim
Page 2: ......
Page 16: ...Section 1 Welcome...
Page 17: ...Chapter 1 About this User Guide...
Page 27: ...Chapter 2 Introduction...
Page 35: ...Chapter 3 t slim Pump Overview...
Page 54: ...Section 2 Key Pump Features...
Page 55: ...Chapter 4 Getting Started...
Page 61: ...Chapter 5 Infusion Site Care and Loading Cartridge...
Page 75: ...Chapter 6 Personal Profile...
Page 85: ...Chapter 7 Bolus...
Page 93: ...Chapter 8 Stop Resume Insulin...
Page 96: ...Section 3 Additional Pump Operations...
Page 97: ...Chapter 9 Temporary Basal Rate...
Page 101: ...Chapter 10 Quick Bolus...
Page 105: ...Chapter 11 t slim Pump Settings...
Page 109: ...Chapter 12 t slim Pump Info and History...
Page 111: ...Chapter 13 t slim Pump Reminders...
Page 117: ...Chapter 14 User Settable Alerts and Alarms...
Page 120: ...Section 4 t slim Pump Safety Alerts and Alarms...
Page 121: ...Chapter 15 t slim Pump Alerts...
Page 143: ...Chapter 16 t slim Pump Alarms...
Page 157: ...Chapter 17 t slim Pump Malfunction...
Page 160: ...Section 5 System Care and Maintenance...
Page 161: ...Chapter 18 Taking Care of Your t slim System...
Page 164: ...Section 6 System Technical Specifications...
Page 165: ...Chapter 19 Technical Specifications...
Page 187: ......